Chapter 3 Operation

Series 24-HP Instruction Manual

 

 

Display Menu

ENTER

Run Mode

Password

ENTER

Display

Menu

Cycle Time(Sec) 0

Number of Digits 2

Display TC(sec) 1

Use keys to access menus

If Cycle Time is set to zero, manual advance is required

Used to set t he number of digits displayed after decimal point

TC = Display Time Constant, used to smooth display

MF Vf Te Pr De T Y or N

A1 A2 A3 Fl Dt E

Y or N

Mf = Mass Flow

Vf = Volume Flow

Te = Temperature

Pr = Pressure

De = Density

T = Total

A1 = Alarm 1 Status

A2 = Alarm 2 Status

A3 = Alarm 3 Status

Fl = Fluid

Dt = Density

* E = Energy

For each parameter:

Select Yes to view parameter in Run Mode Select No to hide parameter in Run Mode

*Energy EMS Meters Only

Use the Display Menu to set the cycle time for automatic screen sequencing used in the Run Mode, change the precision of displayed values, smooth the values or enable or disable each item displayed in the Run Mode screens.

Example for Changing a Run Mode Display Item

The following shows how to remove the temperature screen from the Run Mode screens. Note: all outputs are disabled while using the Setup Menus.

1.Use ￿￿ keys to move to the Display Menu.

2.Press ￿ key until Mf Vf Pr Te De T appears.

3.Press ENTER to select. Press ￿ key until the cursor is positioned below Te.

4.Press ￿ key until N appears. Press ENTER to select.

5.Press EXIT and then ENTER to save changes and return to the Run Mode.

3-6

IM-24-HP

Page 46
Image 46
Sierra 241-VTP, 240-VTP instruction manual Display Menu, Example for Changing a Run Mode Display Item