Series | Chapter 2 Installation |
|
|
Chapter 2 Installation
Installation Overview
Warning! Consult the flow meter name- plate for specific flow meter ap- provals before any hazardous lo- cation installation.
Sierra’s
Flow Meter Installation Requirements
Before installing the flow meter, verify the installation site allows for these considerations:
1.Line pressure and temperature will not exceed the flow meter rating.
2.The location meets the required minimum number of pipe di- ameters upstream and downstream of the sensor head as illus- trated Figure
3.Safe and convenient access with adequate overhead clear- ance for maintenance purposes.
4.Verify that the cable entry into the instrument meets the specific standard required for hazardous area installations.
5.For remote installations, verify the supplied cable length is sufficient to connect the flow meter sensor to the remote electronics.
Also, before installation check your flow system for anomalies such as:
•leaks
•valves or restrictions in the flow path that could create distur- bances in the flow profile that might cause unexpected flow rate indications