Series | Chapter 2 Installation |
|
|
Flange-Style Flow Meter Installation
Caution! When using toxic or corro- sive gases, purge the line with inert gas for a mini- mum of four hours at full gas flow before installing the flow meter.
Install the
Figure 2-4. Flange-Style Flow Meter Installation
When installing the meter make sure the flange marked “inlet” is positioned upstream of the outlet flange, facing the flow. This ensures that the sensor head is positioned downstream of the vortex shedder bar and is correctly aligned to the flow. Installing the meter opposite this direction will result in completely inaccurate flow measurement. To install the meter:
1.Turn off the flow of process gas, liquid or steam. Verify that the line is not pressurized. Confirm that the installation site meets the re- quired minimum upstream and downstream pipe diameters.
2.Seat the meter level and square on the mating connections with the flange marked “inlet” facing the flow. Position a gasket in place for each side. Make sure both gaskets are smooth and even with no gasket material ex- tending into the flow profile. Obstructions in the pipeline will disturb the flow and cause inaccurate measurements.
3.Install bolts in both process connections. Tighten the nuts in the se- quence shown in Figure