Channel Trigger

Trigger modes and appendant registers

 

 

Channel window trigger for entering signals

The analog input is continuously sampled with the selected sample rate. The upper and the lower level define a win- dow. Every time the signal enters the the window from the outside, a triggerevent will be detected.

Register

Value

Direction

set to

Value

SPC_TRIGGERMODE

40000

r/w

TM_CHANNEL

20040

SPC_TRIGGERMODE0

40200

r/w

TM_CHXWINENTER

10040

 

 

 

 

 

SPC_HIGHLEVEL0

42000

r/w

Sets the window’s upper level relatively to the channel’s input range.

board dependant

 

 

 

 

 

SPC_LOWLEVEL0

42100

r/w

Sets the window’s lower level relatively to the channel’s input range.

board dependant

Channel window trigger for leaving signals

The analog input is continuously sampled with the selected sample rate. The upper and the lower level define a win- dow. Every time the signal leaves the the window from the inside, a triggerevent will be detected.

Register

Value

Direction

set to

Value

SPC_TRIGGERMODE

40000

r/w

TM_CHANNEL

20040

SPC_TRIGGERMODE0

40200

r/w

TM_CHXWINLEAVE

10050

SPC_HIGHLEVEL0

42000

r/w

Sets the window’s upper level relatively to the channel’s input range.

board dependant

 

 

 

 

 

SPC_LOWLEVEL0

42100

r/w

Sets the window’s lower level relatively to the channel’s input range.

board dependant

74

MC.31xx Manual

Page 74
Image 74
Spectrum Brands MC.31XX manual Channel window trigger for entering signals, Channel window trigger for leaving signals