Model 9200 Programmable Dual-loop Controller
SSi Manual SERIES 9200 46 Programmable Dual-loop
Pressing the “Next ->” key displays the following calibration screen:
Pressing the “Edit” button displays a numeric keypad allowing you to change the Span input 2 range value.
Normally terminals 29(-) and 30(+) have 17.500 mV applied for this step.
After entering the new Span Input value you must press “Calibrate”. Pressing the “Calibrate” key stores the
appropriate value. Pressing the “Done” will return you to the “User Calibration” Screen.
The current value is displayed directly above the bottom row of buttons as Current Input 2 value: XX.X UV
Edit
Span input 2 range 2 Calibrate
<-- Back Skip Next --> DONE
Enter span voltage (sugg. 65.00 mV)