Model 9200 Programmable Dual-loop Controller

Pressing the “Next ->” key displays the following calibration screen:

Span Output 1

Entered measured output current (mA)

Edit

Calibrate

<-- Back

Skip

Next -->

DONE

Pressing the “Edit” button displays a numeric keypad allowing you to change the Span Output 1 value. Measured at terminals 24(-) and 25(+) for this step. After entering the new Span Output value you must press “Calibrate”. Pressing the “Calibrate” key stores the appropriate value. Pressing the “Done” will return you to the “User Calibration” Screen.

Pressing the “Next ->” key displays the following calibration screen:

Zero Output 2

Enter zero output current (mA)

Edit

Calibrate

<-- Back

Skip

Next -->

DONE

Pressing the “Edit” button displays a numeric keypad allowing you to change the Zero Output 2 value. Measured at terminals 26(-) and 25(+) for this step. After entering the new Zero Output value you must press “Calibrate”. Pressing the “Calibrate” key stores the appropriate value. Pressing the “Done” will return you to the “User Calibration” Screen.