Configuration

 

85

 

 

 

 

 

 

Control Panel Parameters

 

 

 

 

Table 10 • Printer Parameters (Sheet 14 of 27)

 

 

 

 

 

 

 

 

 

 

Parameter

 

Action/Explanation

 

 

 

 

 

 

 

 

 

 

 

Calibrate Media and Ribbon Sensor Sensitivity

 

 

 

 

MEDIA AND RIBBON

 

 

 

 

 

 

Use this procedure to adjust sensitivity of media and ribbon sensors.

 

 

 

CALIBRATE

 

Important • Follow this procedure exactly as presented. All of the

 

 

 

 

 

steps must be performed even if only one of the sensors requires

 

 

 

 

 

adjustment. You may press the left oval at any step in this procedure

 

 

 

 

 

to cancel the process.

 

 

To perform a media and ribbon sensor calibration:

1.Press the right oval to start the calibration procedure. The LOAD BACKING prompt displays.

2.Open the printhead.

3.Remove approximately 8 in. (203 mm) of labels from the backing, and pull the media into the printer so that only the backing is between the media sensors.

4.Leave the printhead open.5.Press the right oval to continue.The REMOVE RIBBON prompt displays.6.Remove the ribbon (if used).7.Close the printhead.8.Press the right oval to continue.The message CALIBRATING PLEASE WAIT displays.

The printer adjusts the scale (gain) of the signals that it receives from the media and ribbon sensors based on the specific media and ribbon combination being used. On the sensor profile, this essentially corresponds to moving the peak of the graph up or down to optimize the readings for your application.

When calibration is complete, RELOAD ALL displays.

9.Open the printhead and pull the media forward until a label is positioned under the media sensor.

10.Reload the ribbon (if used).11.Close the printhead.12.Press the right oval to continue.

The printer performs an auto-calibration. During this process, the printer checks the readings for the media and ribbon based on the new scale established, determines the label length, and determines the print mode. To see the new readings on the new scale, print a sensor profile.