Leica CM3050 S instruction manual Site requirements, Electrical connections

Page 12

2.Safety

2.3Safety instructions: handling and operation of the instrument

2.3.1 Transport

To avoid severe damage to the instrument by running it while the compressor oil is dis- placed from its regular position:

Do not tilt the instrument,

only transport in an upright position.

After transport, wait at least 4 hours be-

fore turning the instrument on.

-->The compressor oil may have been displaced during transport and must settle to its original position be- fore switchingthe instrument on. Otherwise, the instrument may be severely damaged.

2.3.2 Site requirements

Do not operate the instrument in rooms with explosion hazard.

To ensure proper instrument function main- tain a minimum distance of 10 cm between walls and/or furniture and all sides of the in- strument!

2.3.3 Electrical connections

Do not use extension cords for connecting the instrument to mains.

-->Fire hazard!

-->Instrument malfunctions caused by voltage drop.

During the start-up phase of the compressor, the nominal voltage must not drop below the values specified in chapter 4.2 ‘Technical data’!

-->The compressor needs a start-up current at between 25 and 35 A (see chapter 4.2 ‘Technical data’)

Ensure uniform current supply according to specifications.

-->Electrical power supply deviating from specifications damages the instrument.

Therefore, arrange for the electrical in- stallations on site to be checked by a trained professional and make sure any necessary upgrades are installed!

Have the circuit protected by a fuse of its own!

Do not connect any other consumers to the same circuit.

Prior to connecting the instrument to mains, make sure the electrical power supply in your laboratory corresponds to the values indicated on the instrument nameplate (located at the rear of the in- strument)!

12

Instruction Manual V1.2 - 11/2006

Image 12
Contents Leica CM3050 S Page Important note Table of Contents Table of Contents EC Declaration of Conformity Symbols used in this manual and their meaning Important informationSerial- number General Designated useQualification of personnel Any other use of the instrument is considered improperIntegrated safety devices SicherheitSafety Locking the handwheelLocking UnlockingActivating the emergency stop function Deactivating the emergency stopKnife guard Site requirements Electrical connectionsDefrosting/Handling frozen tissue Frozen parts of the instrument and frozen accessoriesHandling microtome knives/blades Knife guard/handwheel lockDisinfection and cleaning Infectious/radioactive material11Removing/reinstalling the microtome MaintenanceDisplay message ‘Dry microtome’ Installation Unpacking and installationGeneral site requirements RepackingStandard delivery Further accessories Inserting the accessories Footswitch Function Models with footswitchConnecting the footswitch Prior to switching on the instrumentInstrument Properties OverviewInstrument Properties Technical Data Microtome Rotary microtome Operation Operating the instrument for the first timeSwitching on/fuses Switching onMains switch / automatic cutout for the instrument Automatic cutout for sectioning motorControl panel Key functions in control panel Display functions in control panelLock / unlock key Activates / deactivates a number of functionsY M I C R O T O M E Specimen temperature Instruments with specimen cooling only 15-minute stepsCryochamber temperature Preset counter Instruments with sectioning motor onlyPress key until menu item ‘Preset counter’ is displayed = Preset counter activatedReal time Starting time of automatic defrost cycleDuration of defrost cycle Press lock/unlock key to activate or de- activate ON/OFFStatus display StopThese parameters can be displayed via the ar- row keys Refrigeration stateSection thickness Total section thickness counter Control panel Instruments without sectioning motor Instruments with sectioning motorActivating a sectioning mode Sectioning mode 1 Intermittent strokeStart / stop motorized sectioning Press sectioning window key once more Emergency stop Functional area B all models Coarse feed keys Coarse feed ‘fast forward’ Travelling speed 1,000 μm/sTo start, press and hold Setting the section thickness Selecting trimming thicknessMotorized trimming ‘Defrosting Handling frozen tissue’ Chapter Daily operation of the instrument‘Knife guard / Handwheel lock’ Chapter Installing the 90 prism Freezing specimens with the 90 prismMobile heat extractor See .3.4 ‘Menu functions’ itemThermoblock Inserting specimen discs into the specimen headSpecimen orientation Trimming / sectioning Inserting the knife Manual trimmingSectioning Insert the knife / disposable blade into the knife holderSwitching off the instrument for a longer period of time Finishing work 5.6.1 Daily routine stepsLock handwheel Defrosting Automatic chamber defrost cycleInstruments with specimen cooling Manual chamber defrost cycle Troubleshooting, applications tips Potential problems causes and remedies Problem Causes RemediesProblem Causes Luke-warm specimen disc Problem Causes Remedies Microtome defectiveReference chart for temperature settings in minus C TissueDisinfection, cleaning and maintenance Safety instructions On disinfection and cleaningSpray disinfection with Leica Cryofect Allow to soak for 15 minutesCleaning Switch the instrument offDisinfection with conventional disinfectants Remove section waste from the chamber and dispose of it‘Disinfection and cleaning’ Chapter Reinstalling the microtome Place plastic coupling 11 on handwheel couplingUnlock handwheel Installing the replacement lamp Exchanging the fluorescent light lampRemoval Removing the sliding windowWarranty and service WarrantyTechnical service information Decommissioning and disposalElectrohydraulic height adjustment Safety instructionsAppendix Installation Maintenance and care TroubleshootingControl keys Ordering information accessories Knife holders AccessoriesSpecimen discs ConsumablesEC Declaration of Conformity Dag Graupner Managing DirectorThis equipment has been cleaned and decontaminated Decontamination Certificate masterDear Customer NB Microtome knives must be in boxesBU Return Authorisation Number