Leica CM3050 S instruction manual Problem Causes

Page 49

6.Troubleshooting, applications tips

Problem

Causes

Remedies

Sections curl on anti-roll plate.

- Anti-roll plate does not pro-

- Readjust correctly.

trude far enough beyond the

 

knife edge.

 

Scraping noise during section- ing stroke and specimen return stroke.

- Anti-roll plate protrudes too far

- Readjust correctly.

beyond the knife edge and is

 

scraping against the specimen.

 

Ondulated sections.

-

Knife damaged.

-

Use different part of the knife

 

 

 

 

or replace knife.

 

-

Edge of anti-roll plate damaged.

-

Replace anti-roll plate.

Chatter marks form during sectioning.

-Specimen insufficiently frozen onto specimen disc.

-Specimen disc not clamped tightly enough.

-Specimen holder ball joint not clamped.

-Knife not clamped tightly enough.

-Selected section thickness too thick - specimen has come off the specimen disc.

-Specimen very hard and inho- mogeneous.

-Blunt knife.

-Knife profile inappropriate for the type of specimen to be sectioned.

-Wrong clearance angle.

-Refreeze specimen onto disc.

-Check disc clamping.

-Check ball joint clamping.

-Check knife clamping.

-Refreeze specimen onto disc.

-Increase section thickness; if possible, reduce specimen surface.

-Use different part of knife or replace knife.

-Use knife with a different pro- file.

-Readjust clearance angle set- ting.

Condensation forms on anti-roll

- Brush, forceps, cloth or other

- Accessories and cleaning

plate and knife during cleaning.

cleaning item too warm.

material to be used pre-cooled

 

 

only. - Store on storage shelves

 

 

inside the cryochamber.

Leica CM3050 S – Cryostat

49

Image 49
Contents Leica CM3050 S Page Important note Table of Contents Table of Contents EC Declaration of Conformity Important information Symbols used in this manual and their meaningSerial- number Designated use GeneralQualification of personnel Any other use of the instrument is considered improperSicherheit Integrated safety devicesLocking the handwheel SafetyLocking UnlockingDeactivating the emergency stop Activating the emergency stop functionKnife guard Electrical connections Site requirementsFrozen parts of the instrument and frozen accessories Defrosting/Handling frozen tissueHandling microtome knives/blades Knife guard/handwheel lockInfectious/radioactive material Disinfection and cleaningMaintenance 11Removing/reinstalling the microtomeDisplay message ‘Dry microtome’ Unpacking and installation InstallationGeneral site requirements RepackingStandard delivery Further accessories Inserting the accessories Models with footswitch Footswitch FunctionConnecting the footswitch Prior to switching on the instrumentOverview Instrument PropertiesInstrument Properties Technical Data Microtome Rotary microtome Operating the instrument for the first time OperationSwitching on Switching on/fusesMains switch / automatic cutout for the instrument Automatic cutout for sectioning motorDisplay functions in control panel Control panel Key functions in control panelLock / unlock key Activates / deactivates a number of functionsY M I C R O T O M E 15-minute steps Specimen temperature Instruments with specimen cooling onlyPreset counter Instruments with sectioning motor only Cryochamber temperaturePress key until menu item ‘Preset counter’ is displayed = Preset counter activatedStarting time of automatic defrost cycle Real timeDuration of defrost cycle Press lock/unlock key to activate or de- activate ON/OFFStop Status displayRefrigeration state These parameters can be displayed via the ar- row keysSection thickness Total section thickness counter Instruments with sectioning motor Control panel Instruments without sectioning motorSectioning mode 1 Intermittent stroke Activating a sectioning modeStart / stop motorized sectioning Press sectioning window key once more Emergency stop Coarse feed ‘fast forward’ Travelling speed 1,000 μm/s Functional area B all models Coarse feed keysTo start, press and hold Selecting trimming thickness Setting the section thicknessMotorized trimming Daily operation of the instrument ‘Defrosting Handling frozen tissue’ Chapter‘Knife guard / Handwheel lock’ Chapter Freezing specimens with the 90 prism Installing the 90 prismMobile heat extractor See .3.4 ‘Menu functions’ itemInserting specimen discs into the specimen head ThermoblockSpecimen orientation Manual trimming Trimming / sectioning Inserting the knifeSectioning Insert the knife / disposable blade into the knife holderFinishing work 5.6.1 Daily routine steps Switching off the instrument for a longer period of timeLock handwheel Automatic chamber defrost cycle DefrostingInstruments with specimen cooling Manual chamber defrost cycle Troubleshooting, applications tips Remedies Potential problems causes and remedies Problem CausesProblem Causes Luke-warm specimen disc Microtome defective Problem Causes RemediesTissue Reference chart for temperature settings in minus CSafety instructions On disinfection and cleaning Disinfection, cleaning and maintenanceSpray disinfection with Leica Cryofect Allow to soak for 15 minutesSwitch the instrument off CleaningDisinfection with conventional disinfectants Remove section waste from the chamber and dispose of it‘Disinfection and cleaning’ Chapter Place plastic coupling 11 on handwheel coupling Reinstalling the microtomeUnlock handwheel Exchanging the fluorescent light lamp Installing the replacement lampRemoval Removing the sliding windowWarranty Warranty and serviceTechnical service information Decommissioning and disposalSafety instructions Electrohydraulic height adjustmentAppendix Installation Troubleshooting Maintenance and careControl keys Accessories Ordering information accessories Knife holdersSpecimen discs ConsumablesDag Graupner Managing Director EC Declaration of ConformityDecontamination Certificate master This equipment has been cleaned and decontaminatedDear Customer NB Microtome knives must be in boxesBU Return Authorisation Number