6.Troubleshooting, applications tips
6.1 Displayed error messages
The autodiagnosis function of the CM3050 S displays error messages in case of certain mal- functions.
1. Power failure
This error message is displayed after the instru- ment had been disconnected from mains, either due to power failure or because the mains switch had been turned off.
C | T | - | 3 | 0 | ° | C |
|
| O | T | - | 3 | 5 | ° | C |
|
| P | O | W | E | R |
| F | A | I | L | U | R | E |
|
•To acknowledge the message, press for about 5 seconds.
2. Coarse feed
Coarse feed malfunction has occurred.
C | T | - | 3 | 0 | ° | C |
|
| O | T | - | 3 | 5 | ° | C |
|
|
| ? | ? | ? | ? |
|
|
|
|
|
|
|
|
|
This message is also displayed, if the coarse feed plug has not been reconnected properly after reinstallation of the microtome.
3. Temperature indications
C T - # # ° C | O T - 3 5 ° C |
or
C | T | - | 3 | 0 | ° | C |
|
| O | T | - | # | # | ° | C |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
If the # symbol is displayed instead of chamber or specimen temperature indication, the corre- sponding temperature sensor is broken or a short circuit has occurred.
4. Service
C T - 3 0 ° C | O T - 3 5 ° C |
S E R V I C E
This message is displayed after a certain number of operating hours to remind the user to have a preven- tive maintenance done by Leica Technical Service.
This message is displayed only, if the instrument is not regularly ser- viced by Leica Technical Service.
We recommend a maintenance contract once the warranty period has expired!
Contact your local Leica sales office for details!
If one of the error messages listed under 2, 3 and 4 is displayed, please contact Leica Technical Service.
6.1.1 Other error indications
(red ) in control panel 2 is illuminated:
•Emergency stop button accidentally activat- ed.
•Footswitch not properly connected.