Leica CM3050 S Daily operation of the instrument, ‘Knife guard / Handwheel lock’ Chapter

Page 40

5.Operation

5.5 Daily operation of the instrument

Select chamber temperature and, in instru- ments with specimen cooling, specimen temperature according to the type of speci- men to be sectioned (see also chapter 6.3 ‘Reference chart for temperature settings’).

Do not continue with step 5.5.1 before the actual chamber / (specimen) temperature corresponds to the set value. This may take several hours (e.g. when operating the in- strument for the first time or when the instru- ment had been switched off).

Prior to operating the instrument , please read and observe the fol- lowing safety instructions:

‘Integrated safety devices’ (Chapter 2.2) .

‘Handling microtome knives

/blades’ (Chapter 2.3.4)

‘Knife guard / Handwheel lock’ (Chapter 2.3.5)

’Motorized sectioning’ (Chapter 2.3.6)

‘Defrosting

/handling frozen tissue’ (Chapter 2.3.7)

‘Frozen instrument parts and accessories’ (Chapter 2.3.8)

‘Infectious / radioactive material’ (Chapter 2.3.9)

5.5.1Freezing specimens onto specimen discs with quick-freeze shelf and stationary heat extractor

Size the specimen.

Apply sufficient amount of embedding medi- um onto ambient temperature specimen disc.

Inside the cryochamber, place the specimen onto the disc and orient.

Place specimen disc with specimen into one of the 10 openings in the quick-freeze shelf (see below) and freeze.

2

 

 

 

1

 

 

 

3

The freezing process can be sped up further by using the stationary heat extractor (1). Next to the low temperature stabilizer (2) (‘Parking station’), 3 specimen discs (3) can be placed within its pivoting range.

Place the heat extractor (1) directly onto the specimen surface. After approx. 30 seconds the specimen is completely frozen.

40

Instruction Manual V1.2 - 11/2006

Image 40
Contents Leica CM3050 S Page Important note Table of Contents Table of Contents EC Declaration of Conformity Important information Symbols used in this manual and their meaningSerial- number General Designated useQualification of personnel Any other use of the instrument is considered improperIntegrated safety devices SicherheitSafety Locking the handwheelLocking UnlockingDeactivating the emergency stop Activating the emergency stop functionKnife guard Site requirements Electrical connectionsDefrosting/Handling frozen tissue Frozen parts of the instrument and frozen accessoriesHandling microtome knives/blades Knife guard/handwheel lockDisinfection and cleaning Infectious/radioactive materialMaintenance 11Removing/reinstalling the microtomeDisplay message ‘Dry microtome’ Installation Unpacking and installationGeneral site requirements RepackingStandard delivery Further accessories Inserting the accessories Footswitch Function Models with footswitchConnecting the footswitch Prior to switching on the instrumentInstrument Properties OverviewInstrument Properties Technical Data Microtome Rotary microtome Operation Operating the instrument for the first timeSwitching on/fuses Switching onMains switch / automatic cutout for the instrument Automatic cutout for sectioning motorControl panel Key functions in control panel Display functions in control panelLock / unlock key Activates / deactivates a number of functionsY M I C R O T O M E Specimen temperature Instruments with specimen cooling only 15-minute stepsCryochamber temperature Preset counter Instruments with sectioning motor onlyPress key until menu item ‘Preset counter’ is displayed = Preset counter activatedReal time Starting time of automatic defrost cycleDuration of defrost cycle Press lock/unlock key to activate or de- activate ON/OFFStatus display StopRefrigeration state These parameters can be displayed via the ar- row keysSection thickness Total section thickness counter Control panel Instruments without sectioning motor Instruments with sectioning motorSectioning mode 1 Intermittent stroke Activating a sectioning modeStart / stop motorized sectioning Press sectioning window key once more Emergency stop Coarse feed ‘fast forward’ Travelling speed 1,000 μm/s Functional area B all models Coarse feed keysTo start, press and hold Selecting trimming thickness Setting the section thicknessMotorized trimming Daily operation of the instrument ‘Defrosting Handling frozen tissue’ Chapter‘Knife guard / Handwheel lock’ Chapter Installing the 90 prism Freezing specimens with the 90 prismMobile heat extractor See .3.4 ‘Menu functions’ itemInserting specimen discs into the specimen head ThermoblockSpecimen orientation Trimming / sectioning Inserting the knife Manual trimmingSectioning Insert the knife / disposable blade into the knife holderFinishing work 5.6.1 Daily routine steps Switching off the instrument for a longer period of timeLock handwheel Automatic chamber defrost cycle DefrostingInstruments with specimen cooling Manual chamber defrost cycle Troubleshooting, applications tips Potential problems causes and remedies Problem Causes RemediesProblem Causes Luke-warm specimen disc Problem Causes Remedies Microtome defectiveReference chart for temperature settings in minus C TissueDisinfection, cleaning and maintenance Safety instructions On disinfection and cleaningSpray disinfection with Leica Cryofect Allow to soak for 15 minutesCleaning Switch the instrument offDisinfection with conventional disinfectants Remove section waste from the chamber and dispose of it‘Disinfection and cleaning’ Chapter Place plastic coupling 11 on handwheel coupling Reinstalling the microtomeUnlock handwheel Installing the replacement lamp Exchanging the fluorescent light lampRemoval Removing the sliding windowWarranty and service WarrantyTechnical service information Decommissioning and disposalSafety instructions Electrohydraulic height adjustmentAppendix Installation Troubleshooting Maintenance and careControl keys Ordering information accessories Knife holders AccessoriesSpecimen discs ConsumablesEC Declaration of Conformity Dag Graupner Managing DirectorThis equipment has been cleaned and decontaminated Decontamination Certificate masterDear Customer NB Microtome knives must be in boxesBU Return Authorisation Number