Leica CM3050 S instruction manual Luke-warm specimen disc

Page 50

6. Troubleshooting, applications tips

Problem

Causes

Remedies

Anti-roll plate damaged after adjustment.

-Anti-roll plate protrudes too far beyond the knife edge. Adjust- ment was done in the direction of the knife.

-Replace anti-roll plate; in future, lift away from the knife when aligning.

-Handle anti-roll plate more carefully.

Thick/thin sections.

-

Wrong temperature for the

-

Select correct temperature.

 

 

tissue to be sectioned.

- Wait until correct temperature

 

 

 

 

has been reached.

 

- Knife profile inappropriate for

- Use knife with different profile

 

 

the specimen to be sectioned.

 

(c or d) or possible switch to

 

 

 

 

disposable blade system.

 

- Ice build-up on the knife back.

-

Remove ice.

 

- Handwheel rotation not uni-

-

Adapt speed.

 

 

form or inappropriate rotation

 

 

 

 

speed.

 

 

 

- Knife not clamped tightly

-

Check knife clamping.

 

 

enough.

 

 

 

-

Specimen disc not clamped

-

Check disc clamping.

 

 

correctly.

 

 

 

- Embedding medium poured

- Place embedding medium on

 

 

onto cold specimen disc; spec-

 

luke-warm specimen disc,

 

 

imen comes off the disc once

 

inside the cryochamber place

 

 

frozen.

 

the specimen onto the disc and

 

 

 

 

freeze in quick-freeze shelf.

 

-

Blunt knife.

- Use different part of knife or

 

 

 

 

replace knife.

 

-

Incorrect clearance angle.

- Set correct clearance angle.

 

-

Specimen dried out.

-

Prepare new specimen.

 

 

 

 

 

Tissue sticks to the anti-roll plate.

-Anti-roll plate too warm or incorrectly adjusted.

-Static electricity.

-Fat on corner or edge of anti- roll plate.

-Knife rusty.

-Cool or readjust anti-roll plate.

-Eliminate cause.

-Remove fat (alcohol).

-Remove rust.

Flattened sections curl up as

-

Static electricity or draft.

-

Eliminate cause.

soon as anti-roll plate is lifted

-

Anti-roll plate too warm.

-

Cool anti-roll plate.

from the knife.

 

 

 

 

50

Instruction Manual V1.1 02/2000

Image 50
Contents Leica CM3050 S Page Important note Table of Contents Table of Contents EC Declaration of Conformity Serial- number Symbols used in this manual and their meaningImportant information Qualification of personnel GeneralDesignated use Any other use of the instrument is considered improperIntegrated safety devices SicherheitLocking SafetyLocking the handwheel UnlockingKnife guard Activating the emergency stop functionDeactivating the emergency stop Site requirements Electrical connectionsHandling microtome knives/blades Defrosting/Handling frozen tissueFrozen parts of the instrument and frozen accessories Knife guard/handwheel lockDisinfection and cleaning Infectious/radioactive materialDisplay message ‘Dry microtome’ 11Removing/reinstalling the microtomeMaintenance General site requirements InstallationUnpacking and installation RepackingStandard delivery Further accessories Inserting the accessories Connecting the footswitch Footswitch FunctionModels with footswitch Prior to switching on the instrumentInstrument Properties OverviewInstrument Properties Technical Data Microtome Rotary microtome Operation Operating the instrument for the first timeMains switch / automatic cutout for the instrument Switching on/fusesSwitching on Automatic cutout for sectioning motorLock / unlock key Control panel Key functions in control panelDisplay functions in control panel Activates / deactivates a number of functionsY M I C R O T O M E Specimen temperature Instruments with specimen cooling only 15-minute stepsPress key until menu item ‘Preset counter’ is displayed Cryochamber temperaturePreset counter Instruments with sectioning motor only = Preset counter activatedDuration of defrost cycle Real timeStarting time of automatic defrost cycle Press lock/unlock key to activate or de- activate ON/OFFStatus display StopSection thickness Total section thickness counter These parameters can be displayed via the ar- row keysRefrigeration state Control panel Instruments without sectioning motor Instruments with sectioning motorStart / stop motorized sectioning Activating a sectioning modeSectioning mode 1 Intermittent stroke Press sectioning window key once more Emergency stop To start, press and hold Functional area B all models Coarse feed keysCoarse feed ‘fast forward’ Travelling speed 1,000 μm/s Motorized trimming Setting the section thicknessSelecting trimming thickness ‘Knife guard / Handwheel lock’ Chapter ‘Defrosting Handling frozen tissue’ ChapterDaily operation of the instrument Mobile heat extractor Installing the 90 prismFreezing specimens with the 90 prism See .3.4 ‘Menu functions’ itemSpecimen orientation ThermoblockInserting specimen discs into the specimen head Sectioning Trimming / sectioning Inserting the knifeManual trimming Insert the knife / disposable blade into the knife holderLock handwheel Switching off the instrument for a longer period of timeFinishing work 5.6.1 Daily routine steps Instruments with specimen cooling DefrostingAutomatic chamber defrost cycle Manual chamber defrost cycle Troubleshooting, applications tips Potential problems causes and remedies Problem Causes RemediesProblem Causes Luke-warm specimen disc Problem Causes Remedies Microtome defectiveReference chart for temperature settings in minus C TissueSpray disinfection with Leica Cryofect Disinfection, cleaning and maintenanceSafety instructions On disinfection and cleaning Allow to soak for 15 minutesDisinfection with conventional disinfectants CleaningSwitch the instrument off Remove section waste from the chamber and dispose of it‘Disinfection and cleaning’ Chapter Unlock handwheel Reinstalling the microtomePlace plastic coupling 11 on handwheel coupling Removal Installing the replacement lampExchanging the fluorescent light lamp Removing the sliding windowTechnical service information Warranty and serviceWarranty Decommissioning and disposalAppendix Electrohydraulic height adjustmentSafety instructions Installation Control keys Maintenance and careTroubleshooting Specimen discs Ordering information accessories Knife holdersAccessories ConsumablesEC Declaration of Conformity Dag Graupner Managing DirectorDear Customer This equipment has been cleaned and decontaminatedDecontamination Certificate master NB Microtome knives must be in boxesBU Return Authorisation Number