5.Operation
5.6Finishing work 5.6.1 Daily routine steps
Remove all specimens from the cry- ochamber and store safely! See also corresponding safety instruc- tions in chapter 2.3.7 ‘Defrosting / handling frozen tissue’!
•Lock handwheel.
•Remove knife from knife holder, disinfect, dry (see chapter 7 ‘Disinfection, cleaning and maintenance’) and store in the knife case.
•Use a cold brush to sweep frozen sectioning waste into the section waste tray.
•Empty section waste tray (dispose of sec- tioning waste according to the applicable regulations on disposal of special category waste!).
•If necessary, carry out a complete disinfection procedure (see chapter 7 ‘Disinfection, cleaning and maintenance’).
•Remove all specimens from the chamber and store safely.
•Close sliding window.
•Switch off cryochamber illumination.
•Use lock/unlock key to lock display in control panel 1 (Display background dark = locked):
C | T | . | . | . | . ° | C |
| O | T | . | . | . | . | ° | C |
|
|
|
|
| 1 | 2 | : | 0 | 0 |
|
|
|
|
|
|
Do not switch off mains switch / automatic cutout (see chapter 5.2 ‘Switching on / fuses’)!
5.6.2Switching off the instrument for a longer period of time
If you do not plan to use the instrument for an extended period of time, it can be switched off.
However, since this involves a lot of work (see below), it should be considered carefully whether switching off the instrument is really the better option.
If the instrument is switched off, the microtome must be removed. Then the microtome, cryo- chamber and all accessories must be disinfect- ed and dried!
•For that purpose, follow the instructions in chapters 7.1 through 7.5.3.
44 | Instruction Manual V1.2 - 11/2006 |