A&D AD-4405 manual To Get Stabilized Data, Specifying the Weighing Range of the First Range

Page 16

Specifying the Weighing Range of the First Range

Step 3 After displaying Cap for 2 seconds, the first range or the weighing capacity of single range will be displayed. When dual range is used, Cap1 is displayed for 2 seconds. Triangle 1 will be displayed. Specify parameter with the 0 - 9 keys and press the ENTER key to store it and proceed to the next step. When pressing the ESC key, regardless of what is displayed, the indicator will proceed to the next step without changing the parameters. The next step is Zero Calibration in single range or the second range resolution in dual range.

Specifying the Second Range Resolution

Step 4 After displaying range2 for 2 seconds, the resolution with decimal point and triangle

2 will be displayed. Specify the second range resolution in the same way as the first range. The decimal point cannot be moved. Specify the second range resolution greater than the first range. Press the ENTER key to store the parameters and proceed to the next step. When pressing the ESC key, regardless of what is displayed, the indicator will proceed to the next step without changing the parameter.

Specifying the Second Range Capacity

Step 5 After displaying Cap2 for 2 seconds, the capacity with unit and decimal point is displayed. Specify the capacity in the same way as the first range. The capacity should be greater than the first range. Press the ENTER key to store the parameters and proceed to Zero Calibration.

5.2.2.To Get Stabilized Data

Step 6 Maintain the following conditions to calibrate the scale (indicator) correctly.

Maintain a constant temperature, stable power and stable input voltage from the load cell.

Avoid direct sunshine or the near the outlet of an air conditioner.

Do not install the scale (indicator) where there is a strong magnetic field.

Step 7 Turn the display on and leave it for several minutes.

5.2. Calibration Procedure

Page 14

AD-4405 Weighing Indicator

Image 16
Contents Weighing Indicator This manual and Marks Contents Setting the Parameters of Setpoint Comparison Compliance with FCC Rules ComplianceCompliance with European Directives Introduction Installation and Precautions Installation and PrecautionsLoad Cell Connections Adjustment of the Load Cell Output Verifying Load Cell Output and Input SensitivityInstalling an Option Board Description of Panels and symbols Front Panel DescriptionOther Displays and Symbols Rear Panel DescriptionAccessories and options Standard accessoriesSupplemental accessories for printer OptionsItems of Calibration Mode CalibrationCalibration Procedure Configuring a Weighing InstrumentSpecifying the Resolution, Decimal Point Position and Format Specifying the Range and UnitSpecifying the Second Range Resolution To Get Stabilized DataSpecifying the Weighing Range of the First Range Specifying the Second Range CapacityZero Calibration Span CalibrationExiting the Calibration Mode Weighing Range Function Setting the Division and Range Digital Linearization FunctionGravity Compensation Function Gravity Acceleration Table Calibration Error Code List Changing the Function Settings FunctionsFunctions OFFZERO+TARE = CAL EXT2 EXT1EXT3 Comparator Data Output Current Loop Output Serial Interface YY/MM/DD MM/DD/YY DD/MM/YYPrint Format CF-Functions Digital Input Preset Tare TareWeighing Tare Clearing TarePreparation and Specifications AccumulationDisplay and Operation Parameter Using Code Memory Code MemoryWeight Check Mode ComparisonHI, OK Condition formula for ComparisonLO, OK Setting Order and Display for Weight Check Mode Setting the Upper/Lower Limit ValuesMode Setpoint Comparison Description of Input parameters and OutputsF22 = 7 Check weighing Output and ConditionF22 = 9 Check weighing 3Loss in weight Simple BatchStart Weighing CompletionToward the Zero band Setting the Parameters of Setpoint ComparisonReturns to Zero band Setting the Hold Functions Hold FunctionDisplay and the Data Output of Hold and Average Relations to the Other FunctionsKey input and command in the hold state Conditions of the Average and ReleaseTARE, ZERO, SETPOINT, Total MT, MZ, HCUnit Weight Registration Using the Counting FunctionCounting Function Key to register the unit weight and return to the piece 13.1 Time and Date Functions Calendar / ClockYY/MM/DD Setting Time and Date Corrected date is as follows 05 05 Internal Printer OP-06 Installing the Printer Paper Special roll paper Replacing the Ink Ribbon15.1 Data Output / Print Mode Common Printer and Data Output ItemsData Number Interval Data Output / PrintSetting the Interval Time Interval Data Output / Print output16.1 Specifications RS-232C InterfaceCircuit and Pin Connection Bit FormatData Format FiguresCommand Format Explanation of CommandsCommands to Request Data Commands to Control the Indicator Changing the Weight Unit Changing the Code MemoryCommands to Set Parameters Set the Preset TareSet the Unit Weight for Counting Mode Set the Limit/Setpoint ValueCommands for the Hold Function Commands to Set the Data Output/ Print Format UFCStart Averaging to Hold Release the Hold DataUFC Commands UFC Command Parameters$DL SF1,$ID$DN$CR$LF$GR$UT$CR$LF’$AN$TL$LF$LF PF2,$DT$TM$LF$AN$SP$TL$LFRS-422/RS-485, Relay OutputOP-03 TRMCOM CircuitSwitching Between RS-422/RS-485 DC50V Relay Output & Control Input OP-05Circuits Settings Specifications19 -20mA Analog Output OP-07 CL1 Current Loop Output OP-08CL2 Current Loop Output Data FormatSpecifications Analog Input and A/D ConversionDigital Section InterfaceDimensions GeneralMemo Memo Memo