Toward the Zero band
Preliminary and Free fall output are holding the off state.
Over/Under comparison starts. If f23 30, judgment starts when Free fall turns on. The judgment is not latched and the output is according to the state at the time. The Weighing completion relay is turned on if f24 2. The on time is set by f25. The READY mark is blinking regardless of the settings of f24 and f25.
Start is may be accepted at this state.
Returns to Zero band
Over/Under and Weighing completion output is off.
The READY mark is turned on.
10.2.3.Setting the Parameters of Setpoint Comparison
Refer to "10.1.2. Setting Upper/Lower Limit Values".
The zero band value is set at f26 of the
Setting Order and Display for Setpoint Comparison and Simple batch
f22 | MODE | Display | 1 | 2 | 3 | 4 | 5 |
7 | Setpoint | Setpoint | Final | Free fall | Over | Under |
|
| Comparison | Class | Fi | FF | Hi | Lo |
|
| 1 | Comparator | OK |
| HI | LO |
|
|
| Triangle |
| 1 |
|
|
|
8 | Setpoint | Setpoint | Free fall | Preliminary | Over | Under | Under |
| Comparison | Class | Fi | FF | Pr | Hi | Lo |
| 2 | Comparator | OK |
|
| HI | LO |
|
| Triangle |
| 1 | 2 |
|
|
9 | Setpoint | Setpoint | Final | Free fall | Full | Over | Under |
| Comparison | Class | Fi | FF | Fu | Hi | Lo |
| 3 | Comparator | OK |
|
| HI | LO |
|
| Triangle |
| 1 | 2 |
|
|
10 | Simple | Setpoint | Final | Free fall | Over | Under |
|
| Batch 1 | Class | Fi | FF | Hi | Lo |
|
|
| Comparator | OK |
| HI | LO |
|
|
| Triangle |
| 1 |
|
|
|
11 | Simple | Setpoint | Final | Free fall | Preliminary | Over | Under |
| Batch 2 | Class | Fi | FF | Pr | Hi | Lo |
|
| Comparator | OK |
|
| HI | LO |
|
| Triangle |
| 1 | 2 |
|
|
12 | Simple | Setpoint | Final | Free fall | Full | Over | Under |
| Batch 3 | Class | Fi | FF | Fu | Hi | Lo |
|
| Comparator | OK |
|
| HI | LO |
|
| Triangle |
| 1 | 2 |
|
|
Page 43 | 10.2. Setpoint Comparison |