A&D AD-4405 manual 19 -20mA Analog Output OP-07, Specifications, Settings

Page 71

19. 4-20mA Analog Output (OP-07)

The OP-07 analog output option is for sending the weight data to an analog input unit. The output is a 4mA to 20mA current output proportional to the display reading.

The output data is updated in synchronization with the display update.

Specifications

Output current

4mA to 20mA *

 

Non-linearity

 

Less than +/- 0.1% fs

Load resistance

0 to 510 W

 

Temperature

ZERO

Less than +/- 0.02% fs/°C

Resolution

Approx. 1/10000

 

coefficient

SPAN

Less than +/- 0.02% fs/°C

Output terminal

Connector terminal

No.1 : + No.2 : - No.3 : FG(Earth)

Accessory

Connector terminal (1) TM-MSTB03STF

 

 

AC adapter (1) Note: Please confirm that the AC adapter type is correct

 

 

 

for your local voltage and receptacle type.

*When set to non-weight display (Calibration, F-settings etc.), output current is 4mA.

The output current is not adjustable.

+ - ￿￿ ￿￿ ￿

 

&%

 

 

￿￿

￿￿

()

 

 

 

 

 

 

 

 

 

DC 7V - 10V

Analog output terminal

 

 

 

 

 

AC adapter Jack

 

 

 

 

 

Settings

Set f30 1 of the F-Function, and set f31, f32 and f33.

 

Item

Parameter

Remarks

 

 

 

 

 

 

 

 

 

 

 

Data

f30

* 0

No output

 

Initial setting

 

 

output

Data output

1

Analog output

 

Must be set to 1

 

 

 

2

Serial in/out 1

 

 

 

 

 

 

3

Serial in/out 2

 

 

 

 

Analog

f31

* 0

Displayed value

 

Initial setting

 

 

output

Output data

1

GROSS weight

 

 

 

 

 

 

2

NET weight

 

 

 

 

 

f32

-999999 to 999999

 

Decimal point

is

set

 

Weight value at 4mA

(Initial setting is 0)

 

at “CALSEt”

 

 

 

f33

-999999 to 999999

 

Decimal point

is

set

 

Weight value at 20mA

(Initial setting is 10000)

 

at “CALSEt”

 

 

Settings of f32 and f33

When entering the f32 or f33 settings (press the

ENTER key when “32” or “33” is blinking), the setting value is displayed. Set the value using the 0 - 9 key. By pressing the +/￿￿ key, the polarity of the value can be alternated. Press

the ENTER key to store the setting value into memory. After this the display returns to selection of the Function number.

AD-4405 Weighing Indicator

Page 69

Output current

20mA

 

b

a

4mA

 

f32 a

f33 a

f33 b

f32 b Weight

19. 4-20mA Analog Output (OP-07)

Image 71
Contents Weighing Indicator This manual and Marks Contents Setting the Parameters of Setpoint Comparison Compliance with European Directives ComplianceCompliance with FCC Rules Introduction Load Cell Connections Installation and PrecautionsInstallation and Precautions Verifying Load Cell Output and Input Sensitivity Adjustment of the Load Cell OutputInstalling an Option Board Front Panel Description Description of Panels and symbolsRear Panel Description Other Displays and SymbolsOptions Accessories and optionsStandard accessories Supplemental accessories for printerCalibration Items of Calibration ModeConfiguring a Weighing Instrument Calibration ProcedureSpecifying the Range and Unit Specifying the Resolution, Decimal Point Position and FormatSpecifying the Second Range Capacity Specifying the Second Range ResolutionTo Get Stabilized Data Specifying the Weighing Range of the First RangeSpan Calibration Zero CalibrationExiting the Calibration Mode Weighing Range Function Digital Linearization Function Setting the Division and RangeGravity Compensation Function Gravity Acceleration Table Calibration Error Code List Functions Changing the Function SettingsOFF FunctionsZERO+TARE = CAL EXT3 EXT1EXT2 Comparator Data Output Current Loop Output Serial Interface DD/MM/YY YY/MM/DD MM/DD/YYPrint Format CF-Functions Clearing Tare Digital Input Preset TareTare Weighing TareAccumulation Preparation and SpecificationsDisplay and Operation Parameter Code Memory Using Code MemoryComparison Weight Check ModeLO, OK Condition formula for ComparisonHI, OK Mode Setting the Upper/Lower Limit ValuesSetting Order and Display for Weight Check Mode Output and Condition Setpoint ComparisonDescription of Input parameters and Outputs F22 = 7 Check weighingWeighing Completion F22 = 9 Check weighing 3Loss in weightSimple Batch StartReturns to Zero band Setting the Parameters of Setpoint ComparisonToward the Zero band Relations to the Other Functions Setting the Hold FunctionsHold Function Display and the Data Output of Hold and AverageMT, MZ, HC Key input and command in the hold stateConditions of the Average and Release TARE, ZERO, SETPOINT, TotalCounting Function Using the Counting FunctionUnit Weight Registration Key to register the unit weight and return to the piece YY/MM/DD Calendar / Clock13.1 Time and Date Functions Setting Time and Date Corrected date is as follows 05 05 Internal Printer OP-06 Replacing the Ink Ribbon Installing the Printer Paper Special roll paperCommon Printer and Data Output Items 15.1 Data Output / Print ModeInterval Data Output / Print Data NumberInterval Data Output / Print output Setting the Interval TimeBit Format 16.1 SpecificationsRS-232C Interface Circuit and Pin ConnectionFigures Data FormatExplanation of Commands Command FormatCommands to Request Data Commands to Control the Indicator Changing the Code Memory Changing the Weight UnitSet the Limit/Setpoint Value Commands to Set ParametersSet the Preset Tare Set the Unit Weight for Counting ModeRelease the Hold Data Commands for the Hold FunctionCommands to Set the Data Output/ Print Format UFC Start Averaging to HoldUFC Command Parameters UFC CommandsPF2,$DT$TM$LF$AN$SP$TL$LF $DLSF1,$ID$DN$CR$LF$GR$UT$CR$LF ’$AN$TL$LF$LFCircuit RS-422/RS-485, Relay OutputOP-03TRM COMSwitching Between RS-422/RS-485 Circuits Relay Output & Control Input OP-05DC50V 19 -20mA Analog Output OP-07 SpecificationsSettings CL2 Current Loop Output OP-08CL1 Data Format Current Loop OutputInterface SpecificationsAnalog Input and A/D Conversion Digital SectionGeneral DimensionsMemo Memo Memo