8. Accumulation

This function accumulates weighing data and stores the total data and the accumulation count. The data is stored in non-volatile memory, and is not lost even if the power is turned off.

8.1.1.Preparation and Specifications

Set the following parameters to use the accumulation function.

Select Cf08 1 for the CF-function so that the accumulation function becomes effective. Specify the method of accumulation and data at f20 of the F-functions.

Specify the inhibit region for accumulation at f21 of the F-functions.

Selection of Accumulation Mode, F20 of the F-functions

There are two methods of accumulation; manual accumulation using the M+ key and automatic accumulation.

The accumulated data can select "positive data only" or "both polarity data". The accumulated data can select "result of comparison is OK only" or "all result".

Accumulation Condition, F21 of the F-functions

In the case of manual accumulation mode, press the M+ key to accumulate weight data when the STABLE mark is displayed.

Data can be accumulated after the weight data enters the "inhibit region for output". When connecting the power cord and turning the display on, the accumulation mode

takes the same action.

Inhibit region for accumulation

f21

 

Description

Add data anytime

f21

0

Stable data can be used anytime

Above

 

5 d

f21

1

Initial setting

 

Above

 

10 d

f21

2

 

 

 

Above

 

20 d

f21

3

 

 

 

Above

 

50 d

f21

4

 

 

 

Caution

Do not set f21 0 for the automatic accumulation mode

If setting f21 0, it may add the same data two times or more.

Limitation of Accumulation Count and Total

The limitation of accumulation count is 999999.

The limitation of total is 999999 ignoring the decimal point.

If exceeding these limitations, the data is not accumulated.

Example: Setting the decimal point to "0.0", the limitation is "99999.9".

8. Accumulation

Page 34

AD-4405 Weighing Indicator

Image 36
Contents Weighing Indicator This manual and Marks Contents Setting the Parameters of Setpoint Comparison Compliance Compliance with FCC RulesCompliance with European Directives Introduction Installation and Precautions Installation and PrecautionsLoad Cell Connections Adjustment of the Load Cell Output Verifying Load Cell Output and Input SensitivityInstalling an Option Board Description of Panels and symbols Front Panel DescriptionOther Displays and Symbols Rear Panel DescriptionAccessories and options Standard accessoriesSupplemental accessories for printer OptionsItems of Calibration Mode CalibrationCalibration Procedure Configuring a Weighing InstrumentSpecifying the Resolution, Decimal Point Position and Format Specifying the Range and UnitSpecifying the Second Range Resolution To Get Stabilized DataSpecifying the Weighing Range of the First Range Specifying the Second Range CapacityZero Calibration Span CalibrationExiting the Calibration Mode Weighing Range Function Setting the Division and Range Digital Linearization FunctionGravity Compensation Function Gravity Acceleration Table Calibration Error Code List Changing the Function Settings FunctionsFunctions OFFZERO+TARE = CAL EXT1 EXT2EXT3 Comparator Data Output Current Loop Output Serial Interface YY/MM/DD MM/DD/YY DD/MM/YYPrint Format CF-Functions Digital Input Preset Tare TareWeighing Tare Clearing TarePreparation and Specifications AccumulationDisplay and Operation Parameter Using Code Memory Code MemoryWeight Check Mode ComparisonCondition formula for Comparison HI, OKLO, OK Setting the Upper/Lower Limit Values Setting Order and Display for Weight Check ModeMode Setpoint Comparison Description of Input parameters and OutputsF22 = 7 Check weighing Output and ConditionF22 = 9 Check weighing 3Loss in weight Simple BatchStart Weighing CompletionSetting the Parameters of Setpoint Comparison Toward the Zero bandReturns to Zero band Setting the Hold Functions Hold FunctionDisplay and the Data Output of Hold and Average Relations to the Other FunctionsKey input and command in the hold state Conditions of the Average and ReleaseTARE, ZERO, SETPOINT, Total MT, MZ, HCUsing the Counting Function Unit Weight RegistrationCounting Function Key to register the unit weight and return to the piece Calendar / Clock 13.1 Time and Date FunctionsYY/MM/DD Setting Time and Date Corrected date is as follows 05 05 Internal Printer OP-06 Installing the Printer Paper Special roll paper Replacing the Ink Ribbon15.1 Data Output / Print Mode Common Printer and Data Output ItemsData Number Interval Data Output / PrintSetting the Interval Time Interval Data Output / Print output16.1 Specifications RS-232C InterfaceCircuit and Pin Connection Bit FormatData Format FiguresCommand Format Explanation of CommandsCommands to Request Data Commands to Control the Indicator Changing the Weight Unit Changing the Code MemoryCommands to Set Parameters Set the Preset TareSet the Unit Weight for Counting Mode Set the Limit/Setpoint ValueCommands for the Hold Function Commands to Set the Data Output/ Print Format UFCStart Averaging to Hold Release the Hold DataUFC Commands UFC Command Parameters$DL SF1,$ID$DN$CR$LF$GR$UT$CR$LF’$AN$TL$LF$LF PF2,$DT$TM$LF$AN$SP$TL$LFRS-422/RS-485, Relay OutputOP-03 TRMCOM CircuitSwitching Between RS-422/RS-485 Relay Output & Control Input OP-05 DC50VCircuits Specifications Settings19 -20mA Analog Output OP-07 Current Loop Output OP-08 CL1CL2 Current Loop Output Data FormatSpecifications Analog Input and A/D ConversionDigital Section InterfaceDimensions GeneralMemo Memo Memo