A&D AD-4405 manual Calibration Error Code List

Page 23

5.6. Calibration Error Code List

Exiting from a Calibration error

ESC key..... The key to return to the point where an error occurred. Retry the operation.

ESC key while pressing the ON/OFF key.

No parameters are changed, CanCel is displayed and the calibration mode is finished. Press the ON/OFF key to turn the display off.

Error Code List

If an error has occurred during the calibration mode, the following code is displayed.

Error code

Description

 

 

err 0

In multi-interval scale. The last division is set to maximum ( d-50 ).

Therefore the next division can not be entered.

 

 

Resolution exceeds 40000. (Resolution = maximum capacity/ minimum

err 1

division)

 

Reduce the maximum capacity or increase the minimum division.

err 2

Load cell output is too large or too small at zero calibration. Check the weighing

unit and load cell. Refer to "3.1.4.Verifying Load Cell Output and Input

err 3

Sensitivity".

 

err 4

Measuring calibration mass, the value exceeded maximum capacity. Reduce

the calibration mass.

 

err 5

The selected calibration mass is smaller than the minimum division.

err 6

The new input sensitivity is less than 0.2 µV/division. Increase the input

sensitivity. Refer to "3.1.4.Verifying Load Cell Output and Input Sensitivity".

 

 

Placing a mass on the weighing unit, the load cell output becomes a negative

err 7

value. Check the load cell cable connections and the direction of load cell

 

mounting.

 

The load cell output exceeds the input range before the maximum capacity.

err 8

Adjust zero balance referring to "3.1.4.Verifying Load Cell Output and Input

Sensitivity". Replace with a load cell designed for a smaller output. Reduce

 

 

maximum capacity.

err 9

The weight value is out of the input range at zero calibration or span

calibration. Check the weighing unit and cables.

 

err 12

The first weighing range is larger than second weighing range.

err 13

An incorrect mass was selected at the digital linearization function.

Select a mass of the following relation. Lnr 1 < Lnr 2 < Lnr 3.

 

g err

An unacceptable value was selected in the gravity acceleration function.

AD-4405 Weighing Indicator

Page 21

5.6. Calibration Error Code List

Image 23
Contents Weighing Indicator This manual and Marks Contents Setting the Parameters of Setpoint Comparison Compliance with European Directives ComplianceCompliance with FCC Rules Introduction Load Cell Connections Installation and PrecautionsInstallation and Precautions Verifying Load Cell Output and Input Sensitivity Adjustment of the Load Cell OutputInstalling an Option Board Front Panel Description Description of Panels and symbolsRear Panel Description Other Displays and SymbolsOptions Accessories and optionsStandard accessories Supplemental accessories for printerCalibration Items of Calibration ModeConfiguring a Weighing Instrument Calibration ProcedureSpecifying the Range and Unit Specifying the Resolution, Decimal Point Position and FormatSpecifying the Second Range Capacity Specifying the Second Range ResolutionTo Get Stabilized Data Specifying the Weighing Range of the First RangeSpan Calibration Zero CalibrationExiting the Calibration Mode Weighing Range Function Digital Linearization Function Setting the Division and RangeGravity Compensation Function Gravity Acceleration Table Calibration Error Code List Functions Changing the Function SettingsOFF FunctionsZERO+TARE = CAL EXT3 EXT1EXT2 Comparator Data Output Current Loop Output Serial Interface DD/MM/YY YY/MM/DD MM/DD/YYPrint Format CF-Functions Clearing Tare Digital Input Preset TareTare Weighing TareAccumulation Preparation and SpecificationsDisplay and Operation Parameter Code Memory Using Code MemoryComparison Weight Check ModeLO, OK Condition formula for ComparisonHI, OK Mode Setting the Upper/Lower Limit ValuesSetting Order and Display for Weight Check Mode Output and Condition Setpoint ComparisonDescription of Input parameters and Outputs F22 = 7 Check weighingWeighing Completion F22 = 9 Check weighing 3Loss in weightSimple Batch StartReturns to Zero band Setting the Parameters of Setpoint ComparisonToward the Zero band Relations to the Other Functions Setting the Hold FunctionsHold Function Display and the Data Output of Hold and AverageMT, MZ, HC Key input and command in the hold stateConditions of the Average and Release TARE, ZERO, SETPOINT, TotalCounting Function Using the Counting FunctionUnit Weight Registration Key to register the unit weight and return to the piece YY/MM/DD Calendar / Clock13.1 Time and Date Functions Setting Time and Date Corrected date is as follows 05 05 Internal Printer OP-06 Replacing the Ink Ribbon Installing the Printer Paper Special roll paperCommon Printer and Data Output Items 15.1 Data Output / Print ModeInterval Data Output / Print Data NumberInterval Data Output / Print output Setting the Interval TimeBit Format 16.1 SpecificationsRS-232C Interface Circuit and Pin ConnectionFigures Data FormatExplanation of Commands Command FormatCommands to Request Data Commands to Control the Indicator Changing the Code Memory Changing the Weight UnitSet the Limit/Setpoint Value Commands to Set ParametersSet the Preset Tare Set the Unit Weight for Counting ModeRelease the Hold Data Commands for the Hold FunctionCommands to Set the Data Output/ Print Format UFC Start Averaging to HoldUFC Command Parameters UFC CommandsPF2,$DT$TM$LF$AN$SP$TL$LF $DLSF1,$ID$DN$CR$LF$GR$UT$CR$LF ’$AN$TL$LF$LFCircuit RS-422/RS-485, Relay OutputOP-03TRM COMSwitching Between RS-422/RS-485 Circuits Relay Output & Control Input OP-05DC50V 19 -20mA Analog Output OP-07 SpecificationsSettings CL2 Current Loop Output OP-08CL1 Data Format Current Loop OutputInterface SpecificationsAnalog Input and A/D Conversion Digital SectionGeneral DimensionsMemo Memo Memo