Page 23
5.6. Calibration Error Code List
Exiting from a Calibration error
ESC key..... The key to return to the point where an error occurred. Retry the operation.
ESC key while pressing the ON/OFF key.
No parameters are changed, CanCel is displayed and the calibration mode is finished. Press the ON/OFF key to turn the display off.
Error Code List
If an error has occurred during the calibration mode, the following code is displayed.
| Error code | Description |
| | |
| err 0 | In multi-interval scale. The last division is set to maximum ( d-50 ). |
| Therefore the next division can not be entered. |
| |
| | Resolution exceeds 40000. (Resolution = maximum capacity/ minimum |
| err 1 | division) |
| | Reduce the maximum capacity or increase the minimum division. |
| err 2 | Load cell output is too large or too small at zero calibration. Check the weighing |
| unit and load cell. Refer to "3.1.4.Verifying Load Cell Output and Input |
| err 3 |
| Sensitivity". |
| |
| err 4 | Measuring calibration mass, the value exceeded maximum capacity. Reduce |
| the calibration mass. |
| |
| err 5 | The selected calibration mass is smaller than the minimum division. |
| err 6 | The new input sensitivity is less than 0.2 µV/division. Increase the input |
| sensitivity. Refer to "3.1.4.Verifying Load Cell Output and Input Sensitivity". |
| |
| | Placing a mass on the weighing unit, the load cell output becomes a negative |
| err 7 | value. Check the load cell cable connections and the direction of load cell |
| | mounting. |
| | The load cell output exceeds the input range before the maximum capacity. |
| err 8 | Adjust zero balance referring to "3.1.4.Verifying Load Cell Output and Input |
| Sensitivity". Replace with a load cell designed for a smaller output. Reduce |
| |
| | maximum capacity. |
| err 9 | The weight value is out of the input range at zero calibration or span |
| calibration. Check the weighing unit and cables. |
| |
| err 12 | The first weighing range is larger than second weighing range. |
| err 13 | An incorrect mass was selected at the digital linearization function. |
| Select a mass of the following relation. Lnr 1 < Lnr 2 < Lnr 3. |
| |
| g err | An unacceptable value was selected in the gravity acceleration function. |
AD-4405 Weighing Indicator | Page 21 | 5.6. Calibration Error Code List |
Contents
Weighing Indicator
This manual and Marks
Contents
Setting the Parameters of Setpoint Comparison
Compliance with European Directives
Compliance
Compliance with FCC Rules
Introduction
Load Cell Connections
Installation and Precautions
Installation and Precautions
Verifying Load Cell Output and Input Sensitivity
Adjustment of the Load Cell Output
Installing an Option Board
Front Panel Description
Description of Panels and symbols
Rear Panel Description
Other Displays and Symbols
Options
Accessories and options
Standard accessories
Supplemental accessories for printer
Calibration
Items of Calibration Mode
Configuring a Weighing Instrument
Calibration Procedure
Specifying the Range and Unit
Specifying the Resolution, Decimal Point Position and Format
Specifying the Second Range Capacity
Specifying the Second Range Resolution
To Get Stabilized Data
Specifying the Weighing Range of the First Range
Span Calibration
Zero Calibration
Exiting the Calibration Mode
Weighing Range Function
Digital Linearization Function
Setting the Division and Range
Gravity Compensation Function
Gravity Acceleration Table
Calibration Error Code List
Functions
Changing the Function Settings
OFF
Functions
ZERO+TARE = CAL
EXT3
EXT1
EXT2
Comparator
Data Output
Current Loop Output
Serial Interface
DD/MM/YY
YY/MM/DD MM/DD/YY
Print Format
CF-Functions
Clearing Tare
Digital Input Preset Tare
Tare
Weighing Tare
Accumulation
Preparation and Specifications
Display and Operation
Parameter
Code Memory
Using Code Memory
Comparison
Weight Check Mode
LO, OK
Condition formula for Comparison
HI, OK
Mode
Setting the Upper/Lower Limit Values
Setting Order and Display for Weight Check Mode
Output and Condition
Setpoint Comparison
Description of Input parameters and Outputs
F22 = 7 Check weighing
Weighing Completion
F22 = 9 Check weighing 3Loss in weight
Simple Batch
Start
Returns to Zero band
Setting the Parameters of Setpoint Comparison
Toward the Zero band
Relations to the Other Functions
Setting the Hold Functions
Hold Function
Display and the Data Output of Hold and Average
MT, MZ, HC
Key input and command in the hold state
Conditions of the Average and Release
TARE, ZERO, SETPOINT, Total
Counting Function
Using the Counting Function
Unit Weight Registration
Key to register the unit weight and return to the piece
YY/MM/DD
Calendar / Clock
13.1 Time and Date Functions
Setting Time and Date
Corrected date is as follows 05 05
Internal Printer OP-06
Replacing the Ink Ribbon
Installing the Printer Paper Special roll paper
Common Printer and Data Output Items
15.1 Data Output / Print Mode
Interval Data Output / Print
Data Number
Interval Data Output / Print output
Setting the Interval Time
Bit Format
16.1 Specifications
RS-232C Interface
Circuit and Pin Connection
Figures
Data Format
Explanation of Commands
Command Format
Commands to Request Data
Commands to Control the Indicator
Changing the Code Memory
Changing the Weight Unit
Set the Limit/Setpoint Value
Commands to Set Parameters
Set the Preset Tare
Set the Unit Weight for Counting Mode
Release the Hold Data
Commands for the Hold Function
Commands to Set the Data Output/ Print Format UFC
Start Averaging to Hold
UFC Command Parameters
UFC Commands
PF2,$DT$TM$LF$AN$SP$TL$LF
$DL
SF1,$ID$DN$CR$LF$GR$UT$CR$LF
’$AN$TL$LF$LF
Circuit
RS-422/RS-485, Relay OutputOP-03
TRM
COM
Switching Between RS-422/RS-485
Circuits
Relay Output & Control Input OP-05
DC50V
19 -20mA Analog Output OP-07
Specifications
Settings
CL2
Current Loop Output OP-08
CL1
Data Format
Current Loop Output
Interface
Specifications
Analog Input and A/D Conversion
Digital Section
General
Dimensions
Memo
Memo
Memo