A&D AD-4405 manual Setting Time and Date

Page 51

f53 2 Add 2 digit alphabetical characters including space or dot. There are 3 types, a character and a space, a character and a dot, and one space and a character. Setting specifies 2 digit numbers from the next table. If the set number is excluded from the table at f54, two spaces are output.

 

 

Output characters and f54 setting

(f53 2)

(_ : space)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Out

f54

Out

f54

Out

f54

Out

f54

 

Out

f54

Out

f54

A_

01

N_

14

A.

31

N.

44

 

_A

61

_N

74

B_

02

O_

15

B.

32

O.

45

 

_B

62

_O

75

C_

03

P_

16

C.

33

P.

46

 

_C

63

_P

76

D_

04

Q_

17

D.

34

Q.

47

 

_D

64

_Q

77

E_

05

R_

18

E.

35

R.

48

 

_E

65

_R

78

F_

06

S_

19

F.

36

S.

49

 

_F

66

_S

79

G_

07

T_

20

G.

37

T.

50

 

_G

67

_T

80

H_

08

U_

21

H.

38

U.

51

 

_H

68

_U

81

I_

09

V_

22

I.

39

V.

52

 

_I

69

_V

82

J_

10

W_

23

J.

40

W.

53

 

_J

70

_W

83

K_

11

X_

24

K.

41

X.

54

 

_K

71

_X

84

L_

12

Y_

25

L.

42

Y.

55

 

_L

72

_Y

85

M_

13

Z_

26

M.

43

Z.

56

 

_M

73

_Z

86

Example: Set f54 38, internal year (after f52 calculation) = 17, output year = H.17

13.2. Setting Time and Date

Step 1 By pressing the PT(4) key while pressing the PRINT key, time will be displayed.

Use the following keys.

MODE

.......key

The key to select time( time ) or date( date ) to set.

 

 

 

The key to return to the weight display.

ESC

..........key

 

 

The key to proceed to step 2 ( time ) or step 3 ( date ).

ENTER

......key

Step 2 The present time is displayed with f50 setting and is updating. A triangle indicates

AM (triangle 2) or PM (triangle 3) and no triangle is displayed in 24 hour setting. Set the time using the following keys. Once an effective key is entered, the time

stops updating.

 

 

 

 

 

0 - 9 key

The key to set the value at the blinking digit, after that the

 

 

 

 

blinking digit shift to the right. (Illegal numbers are ignored. )

+/- key

The key to shift the blinking digit to the right (including the triangle).

 

 

 

key

The key to reset the settings and displays the present time.

CLEAR

 

 

 

 

The key to change AM and PM when triangle 2 or 3 is blinking.

MODE

.......key

 

 

 

 

 

 

 

 

ESC

..........key

The key to return to

time

display without changing the time

 

 

 

 

settings.

 

 

 

 

 

ENTER

key

The key to store the time then return to the

time

display.

AD-4405 Weighing Indicator

Page 49

13. Calendar / Clock

Image 51
Contents Weighing Indicator This manual and Marks Contents Setting the Parameters of Setpoint Comparison Compliance Compliance with FCC RulesCompliance with European Directives Introduction Installation and Precautions Installation and PrecautionsLoad Cell Connections Verifying Load Cell Output and Input Sensitivity Adjustment of the Load Cell OutputInstalling an Option Board Front Panel Description Description of Panels and symbolsRear Panel Description Other Displays and SymbolsOptions Accessories and optionsStandard accessories Supplemental accessories for printerCalibration Items of Calibration ModeConfiguring a Weighing Instrument Calibration ProcedureSpecifying the Range and Unit Specifying the Resolution, Decimal Point Position and FormatSpecifying the Second Range Capacity Specifying the Second Range ResolutionTo Get Stabilized Data Specifying the Weighing Range of the First RangeSpan Calibration Zero CalibrationExiting the Calibration Mode Weighing Range Function Digital Linearization Function Setting the Division and RangeGravity Compensation Function Gravity Acceleration Table Calibration Error Code List Functions Changing the Function SettingsOFF FunctionsZERO+TARE = CAL EXT1 EXT2EXT3 Comparator Data Output Current Loop Output Serial Interface DD/MM/YY YY/MM/DD MM/DD/YYPrint Format CF-Functions Clearing Tare Digital Input Preset TareTare Weighing TareAccumulation Preparation and SpecificationsDisplay and Operation Parameter Code Memory Using Code MemoryComparison Weight Check ModeCondition formula for Comparison HI, OKLO, OK Setting the Upper/Lower Limit Values Setting Order and Display for Weight Check ModeMode Output and Condition Setpoint ComparisonDescription of Input parameters and Outputs F22 = 7 Check weighingWeighing Completion F22 = 9 Check weighing 3Loss in weightSimple Batch StartSetting the Parameters of Setpoint Comparison Toward the Zero bandReturns to Zero band Relations to the Other Functions Setting the Hold FunctionsHold Function Display and the Data Output of Hold and AverageMT, MZ, HC Key input and command in the hold stateConditions of the Average and Release TARE, ZERO, SETPOINT, TotalUsing the Counting Function Unit Weight RegistrationCounting Function Key to register the unit weight and return to the piece Calendar / Clock 13.1 Time and Date FunctionsYY/MM/DD Setting Time and Date Corrected date is as follows 05 05 Internal Printer OP-06 Replacing the Ink Ribbon Installing the Printer Paper Special roll paperCommon Printer and Data Output Items 15.1 Data Output / Print ModeInterval Data Output / Print Data NumberInterval Data Output / Print output Setting the Interval TimeBit Format 16.1 SpecificationsRS-232C Interface Circuit and Pin ConnectionFigures Data FormatExplanation of Commands Command FormatCommands to Request Data Commands to Control the Indicator Changing the Code Memory Changing the Weight UnitSet the Limit/Setpoint Value Commands to Set ParametersSet the Preset Tare Set the Unit Weight for Counting ModeRelease the Hold Data Commands for the Hold FunctionCommands to Set the Data Output/ Print Format UFC Start Averaging to HoldUFC Command Parameters UFC CommandsPF2,$DT$TM$LF$AN$SP$TL$LF $DLSF1,$ID$DN$CR$LF$GR$UT$CR$LF ’$AN$TL$LF$LFCircuit RS-422/RS-485, Relay OutputOP-03TRM COMSwitching Between RS-422/RS-485 Relay Output & Control Input OP-05 DC50VCircuits Specifications Settings19 -20mA Analog Output OP-07 Current Loop Output OP-08 CL1CL2 Data Format Current Loop OutputInterface SpecificationsAnalog Input and A/D Conversion Digital SectionGeneral DimensionsMemo Memo Memo