Page 56

Description of "Command Mode" (For Serial Interface)

Function

f40 5

Operation

The serial data can be output by the serial interface command only. Data

 

conflicts or accidental data output by mis-operation can be avoided when

 

using multiple instruments with the RS-422/485 interface.

Description of "Delay for continual data"

Function

f38, f42

Operation

This function can be used in the "Auto print mode" and "Manual print mode".

 

When using a non-buffered printer, set to f42 3 and f43 1 (or f38 1).

15.2. Data Number

The data number increments automatically with each serial data output or print out respectively from 1 to 99999 (the next of 99999 is 1). The starting number can be set using the keys.

Setting the Serial Data Number

Press the CODE(1) key while pressing the PRINT key at the weight display, the display changes to data r then r xxx. “xxx” is the next data number. Use the 0 - 9 keys to enter the number, then press the ENTER key to store the number and return to the weight display.

Setting the Print Data Number

Press the SAMPLE(2) key while pressing the PRINT key at the weight display, the display changes to data p then p xxx. “xxx” is the next data number. Use the 0 - 9 keys to enter the number, then press the ENTER key to store the number and return to the weight display.

Note 1. At the number setting, the following keys are available.

+/- key

Sets the number to 1.

 

 

 

Sets the number to 0.

ZERO

key

 

 

Returns to the weight display without changing the number.

ESC

key

2. When setting the number to 0, it is entered as 1 automatically.

15.3. Interval Data Output / Print

The AD-4405 can output and/or print data with a constant time interval using the internal calendar/clock function. The time intervals for serial data output and for the printer can be set individually from 1 second up to 23 hours and 59 minutes.

Caution The interval time must be set to allow enough time to finish output or printing out.

15. Common Printer and Data Output

Page 54

AD-4405 Weighing Indicator

Image 56
Contents Weighing Indicator This manual and Marks Contents Setting the Parameters of Setpoint Comparison Compliance with European Directives ComplianceCompliance with FCC Rules Introduction Load Cell Connections Installation and PrecautionsInstallation and Precautions Adjustment of the Load Cell Output Verifying Load Cell Output and Input SensitivityInstalling an Option Board Description of Panels and symbols Front Panel DescriptionOther Displays and Symbols Rear Panel DescriptionAccessories and options Standard accessoriesSupplemental accessories for printer OptionsItems of Calibration Mode CalibrationCalibration Procedure Configuring a Weighing InstrumentSpecifying the Resolution, Decimal Point Position and Format Specifying the Range and UnitSpecifying the Second Range Resolution To Get Stabilized DataSpecifying the Weighing Range of the First Range Specifying the Second Range CapacityZero Calibration Span CalibrationExiting the Calibration Mode Weighing Range Function Setting the Division and Range Digital Linearization FunctionGravity Compensation Function Gravity Acceleration Table Calibration Error Code List Changing the Function Settings FunctionsFunctions OFFZERO+TARE = CAL EXT3 EXT1EXT2 Comparator Data Output Current Loop Output Serial Interface YY/MM/DD MM/DD/YY DD/MM/YYPrint Format CF-Functions Digital Input Preset Tare TareWeighing Tare Clearing TarePreparation and Specifications AccumulationDisplay and Operation Parameter Using Code Memory Code MemoryWeight Check Mode ComparisonLO, OK Condition formula for ComparisonHI, OK Mode Setting the Upper/Lower Limit ValuesSetting Order and Display for Weight Check Mode Setpoint Comparison Description of Input parameters and OutputsF22 = 7 Check weighing Output and ConditionF22 = 9 Check weighing 3Loss in weight Simple BatchStart Weighing CompletionReturns to Zero band Setting the Parameters of Setpoint ComparisonToward the Zero band Setting the Hold Functions Hold FunctionDisplay and the Data Output of Hold and Average Relations to the Other FunctionsKey input and command in the hold state Conditions of the Average and ReleaseTARE, ZERO, SETPOINT, Total MT, MZ, HCCounting Function Using the Counting FunctionUnit Weight Registration Key to register the unit weight and return to the piece YY/MM/DD Calendar / Clock13.1 Time and Date Functions Setting Time and Date Corrected date is as follows 05 05 Internal Printer OP-06 Installing the Printer Paper Special roll paper Replacing the Ink Ribbon15.1 Data Output / Print Mode Common Printer and Data Output ItemsData Number Interval Data Output / PrintSetting the Interval Time Interval Data Output / Print output16.1 Specifications RS-232C InterfaceCircuit and Pin Connection Bit FormatData Format FiguresCommand Format Explanation of CommandsCommands to Request Data Commands to Control the Indicator Changing the Weight Unit Changing the Code MemoryCommands to Set Parameters Set the Preset TareSet the Unit Weight for Counting Mode Set the Limit/Setpoint ValueCommands for the Hold Function Commands to Set the Data Output/ Print Format UFCStart Averaging to Hold Release the Hold DataUFC Commands UFC Command Parameters$DL SF1,$ID$DN$CR$LF$GR$UT$CR$LF’$AN$TL$LF$LF PF2,$DT$TM$LF$AN$SP$TL$LFRS-422/RS-485, Relay OutputOP-03 TRMCOM CircuitSwitching Between RS-422/RS-485 Circuits Relay Output & Control Input OP-05DC50V 19 -20mA Analog Output OP-07 SpecificationsSettings CL2 Current Loop Output OP-08CL1 Current Loop Output Data FormatSpecifications Analog Input and A/D ConversionDigital Section InterfaceDimensions GeneralMemo Memo Memo