Return to Section TOC

Section TOC

Return to Master TOC

Master TOC

F-41

TROUBLESHOOTING AND REPAIR

F-41

 

AUXILIARY TRANSFORMER TEST (continued)

 

 

 

 

TABLE F.6

 

 

 

 

 

 

 

OUTPUT LOAD CURRENT

 

EXPECTED TRANSDUCER FEEDBACK

 

 

VOLTAGE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

300

 

2.4

 

 

 

250

 

2.0

 

 

 

200

 

1.6

 

 

 

150

 

1.2

 

 

 

100

 

0.8

 

 

 

 

 

 

 

 

Return to

Return to Section TOC

Return to

Return to Master TOC

8.Check the feedback voltage from the current transducer using a resistive load bank and with the POWER WAVE 355M/405M in mode

200.Mode 200 is a constant current test mode. This mode can be accessed using a wire feeder placed in mode 200 or a laptop computer and the appropriate software. Apply the grid load across the output of the POWER WAVE 355M/405M. Set machine output to

300amps and enable WELD TERMINALS. Adjust the grid load to obtain 300 amps on the external ammeter and check feedback volt- ages per Table F.6.

A.Pin 1 (lead 801) to Pin 6 (lead 806) should read 2.4 VDC (machine loaded to 300 amps).

9.If for any reason the machine cannot be loaded to 300 amps, Table F.6. shows what feedback voltage is produced at various cur- rent loads.

10.If the correct supply voltages are applied to the current transducer, and with the machine loaded, the feedback voltage is missing or not correct the current transducer may be faulty.

Before replacing the current transducer, check the leads and plugs between the control board (J8) and the current transducer (J90). See The Wiring Diagram. For access to plug J90 and the current transducer refer to: Current

Transducer Removal and Replacement Procedure.

11.Remove input power to the POWER WAVE 355M/405M.

12.Replace the control box top and any cable ties previously removed.

13.Install the case wraparound cover using the 5/16” nut driver.

Return to Section TOC

Return to Master TOC

POWER WAVE 355M/405M

Page 87
Image 87
Lincoln Electric 355M, 405M service manual 300 250 200 150 100, Transducer Removal and Replacement Procedure