pi'

(,'_,._,

~ . ~' .

·,·_1· !

,1,:.:

R-4106.A

2. Binocular Body

2-6. Levei difference of paired exJt pupils

1)lf in 2-5, 3), the level diiference is found over 0.5mm, ra1se or lower ·§in the direction of the arrow, as shown in Fig. 37.

r;, \\ /

(/(

 

 

 

(-)

(+)

 

Fig. 37

2)After the adjustment oi has been completed, check at ali times for the correct adjust· ments mentioned in 2-3, 2-4and 2-5.

2._ 7. Finally, lóck prisms @ and @ in position using adhesive.

t:.... . e

2-8. Focus adjustment of Photo·tube (UW-tube)

1)lnsert the parfocality adjusting mirror (tool) into photo·tube sleeve @.

2)Setting the autocollimator to infinity, perform infinity checking, using the above mentioned

mirror.

3)If not adjusted to infinity, tum @ in relaiion to photo-tube G).

4) Apply screw lock to G) and @), and leave them in th~'ir adjusted positions until the lock sets, to avoid parfocality derangement caused by screw slack.

-47 -

Page 117
Image 117
Nikon 50i and E100 manual Binocular Body