R-4106.A

AOJUSTMENT

2. Binocular Body

 

 

(ER·tub!DI

1)Unscrew @. and remove the binocular body. Release screw @ to move lens chamber

G)(with ~ and inside). Attach the binocular body to the main body again holding

this with the hand. ·

See that no derangement takes place (For adjustment, consider such a derangement as caused by fastening (@).

2)Apply No. 350 to screw @.

3)Fix the binocular body to the main body Q) with screws x3.

IF-tuoo)

1)Fasten screw e@ lightly, and move lens charnber @) (with @:D and @ inside ). Attach the binocular body to the main body again holding this with the hand to see that no derangement takes place.

2)Lock @ (@), applying No. 350.

3)Fix the binocular body to the main body (D with screws @ x 3.

(UW-tuool

1)Place the straight tube (too!) on t.he binocular tube attaching part, pushing the too! tube sufficiently against the main body. Make adjustment by moving @ (with @ and

@metais inside) within a range of 2 scale lines in the Kellner eyepiece. (If an attaching errar of optical axis of binocular body is found, make adjustment for @, once again.)

2-5. Parfocality adjustment

1)After focusing in the standard straight tube, take reading of the dia! gauge to measure the tube length errar (115 .2 ± 1mm), using too! 125148 for adjusting tube length.

2)Moving vertically the sleeve cover ring e@ (@), bring the above mentioned errar within ±0.lmm.

3)If in the above case, levei difference of paired exit pupils is found within 0.5mm between the right and left ends, OK.

-46-