REPAIR ANO AOJUSTMENT

4. Wobbling in X {right and left) Movement

 

 

4.Wobbling in X (right and left) Movement

Cause: (!) Slack, sticking, squeezing in the X gu1de rail assembly.

a)Inferior meshing of the rack @ and the pinion @.

b)Dent on the X guide rails @ , @ ,the rollers@ or deformed retainers

(j).

e)Dust, soil, scratch on teeth of the rack @ or the pinion @.

(2)Looseness of the screws @ or inferior adjustment of the X guide block@.

<.Repair and Adjustment >

(!)-a) If the rack @ and pinion @ rneshing is inferior adjust the position of the rack and pirúon.

Procedure

i)Detach the slide holder assembly from the stage.

ü) Releasing the screws@x4 separate the X guide block@from the Y guide

bloclcs @ , @ .

Then, the X movement block(D, the rollers@, the retainers(Z) and the guide

rails @ , @ will be rnade free simultaneously.

üi) The rack @ position is to be adjusted slightly releasing the screws @x 2 .

iv)The pinion @ position is to be adjusted slightly releasing the screws@ • 2.

v)Follow the reassembling procedure.

vi)Check the rack and pinion rneshing and the wobbling in the X rnovement.