REASSEMBLY PROCEOURE

2. Substage Unit

 

 

6)To attach the condenser mount follow the procedure as below (Fig. 27):

i)Attach spring retainer e@ to condenser receptacle @) with screws @ . 2.

ii)Insert plate spring @ between @ and Q8) . Bend the spnng as shown by

the arrow in Fig. 39.

ili) Apply G79 IO to the sliding surface of condenser

centering mount @ , and condenser receptacle

@.

iv) Insert @ into @.

Fig. 39

v)Screw condenser clamp screw @ into @ through the opening of@ .

vi)See that no seizure occurs between @ and @ , when manipulating right and left condenser centering screws @ to bring @ to the center, and @ operates smoo thly.