R-4104 A
REASSEMBLY PROCEOURE | 2. Substage Unit |
|
|
6)To attach the condenser mount follow the procedure as below (Fig. 27):
i)Attach spring retainer e@ to condenser receptacle @) with screws @ . 2.
ii)Insert plate spring @ between @ and Q8) . Bend the spnng as shown by
the arrow in Fig. 39.
ili) Apply G79 IO to the sliding surface of condenser
centering mount @ , and condenser receptacle
@.
iv) Insert @ into @. | Fig. 39 |
v)Screw condenser clamp screw @ into @ through the opening of@ .
vi)See that no seizure occurs between @ and @ , when manipulating right and left condenser centering screws @ to bring @ to the center, and @ operates smoo thly.