R-4108.A

 

REPAIR ANO ADJUSTMENT

4. Wobbling in X (right and left) Movement

 

5. Wobbling in Y (to and fro) Mov11m11nt

 

 

 

 

 

 

(2)Loosened screws @ are to be tightened firm and inferior position of (D is to be readjusted.

Procedure /

Slightly releasing the screws @x 2 adjust the G position and check wobbling in the

Xmovement.

5.Wobbling in Y (to and fro) Movement

Cause: ( l) Slack, sticking, squeezing in the Y guide rai! assembly.

a)Inferior meshing of the rack @ and the pinion @ .

b)Dent on the Y guide rails @ , the rollers@or deformed retainers

G).

e)Dust, soil or scratch on teeth of the rack @ or the pinion @.

(2)Looseness of the screws or inferior adjustment of the Y guide block

@(or@).

<Rapar and Adjustment >

(1 )-a) If the rack@ and pinion @ rneshing is inferior adjust their position.

Procedure

i)The rack @ position is to be adjusted slightly releasing the screws @ •2.

ü) Check wobbling in the Y rnovement.

- 15 -

Page 64
Image 64
Nikon 50i and E100 manual Procedure, Repair ANO Adjustment, Wobbling in Y to and fro Movement, Rapar and Adjustment