R-4104.A

REASSEMBL Y PROCEDURE

1. Coaxial Coarse and Fine Focus Knob Unit

 

 

REASSEMBLY PROCEDURE

1.Coaxial Coarse and Fine Focus Knob Unit

1)Beforehand, clean@, e@) , @, @ - @, @. @ usmg tridane.

2)Applying No. 616 to the thread of pinion shaft@, screw it into spur gear@.

3)Apply G7920 to @ at the position indicated in Fig. 16, and insert the subassembly consisting of@ and @ into pinion case @. Furthermore, attach pinion thrust receptacle e@ to @. [n this case, make sure of smooth rotation of @ and

@.

4)Clean the thread of pinion case @, using gasoline.

After drying, apply No. 350, and screw in coarse focus beanng @.(Fig. 36).

(

616) Lubricant (G7920)

--------'""--'-~~

1

~

Fig. 36

5)Insert the above subassembly into the arm. (Fig. 37).

6)Inserting rack (V from undemeath the arm, attach it to the bali race by means of PM screw @ and CM screw @>.(Fig. 38). After attaching, fi.x pinion case @ to the arm by means of screws ~ xJ. (Fig. 37).

~ote: e@ might be scratched or deformed by the use of a wrench, when detached, in this case, replace it.

- 32 -

Page 35
Image 35
Nikon 50i and E100 manual Reassembl Y Procedure, Coaxial Coarse and Fine Focus Knob Unit, Reassembly Procedure