R-4104.A

 

ADJUSTMENT

5. Cantllring the Tube Attmching and Stago Attaching Paru

 

6.

Cantering of OptiCllll System in the Microscope Base

 

 

 

 

 

 

 

5-3. Using positioning tool 125141, to check the position of substage and condenser mount

1)As shown in Fig. 1 (P. 3), bring the part@, of the tool onto the condenser attaching surface, and the part@ onto the stage attaching surface.

2)Raise the condenser attaching surface to

make sure that the part @ enters the part @.

3)If it does not, make adjustment by means of @ x4.

Fig. 47

6. Cantering of Optical System in the Microscope Base

6-l. Using tools 125146 (for OPTIPHOT)

1)Take off the lamp housing, and set the too! 125146. (Fig. 48).

2)Set the above cited collimators on the specified positions on the base.

Elimina te any slack, if takes place, by means of

 

 

125146

 

washers.

 

,.·

3) Bringing the part @CFig. 6) onto the position

Fig. 48

 

 

without the lens, make sure of the correct

positions of the filament (G)) and mirrar (@), viewing through the eyepiece of collima tor. (Fig. 49 ).

4)If the positions are not found coincided, make adjustment by means of screws

@x:i, provided for attaching mirror mount @.(Fig. 32).

5)P1ace the part @onto the positfon with the lens.

Viewing the position of the field diaphragm through the eyepiece, bring the diaphragm to the center by means of screws @ x 2 provided for attaching biade mount @.

~. ;

-41 -

Page 44
Image 44