R-4104.A
DISASSEMBLY PROCEDURE | 2. Subrtage Unit |
|
|
2.Subrtage Unit (OPTIPHOT, LABOPHOT) (Fig. 11. 12)
l)To remove the substage as a unit, release clamp screw @ .
(LABOPHOT: First. remove rear eover 0 and substage body @ from rotler race (male)
(j)).
2)After releasing two set screws @ on the bottom of @, remove pinion bearing
@.(Fig. 28).
3)Pinion ª and pinion shait @ can be taken off, after condenser focus knob @ (combined with bush @) has been removed. (Fig. 28).
4)After unscrewing PM screw '®on the @ , remove the condenser carrier subassembly. (Fig. 28).
Condenser centering mount @ and dovetail @ having been attached in the correct position, using the tool, it is not recommended to detach them, so far as no problem arises. (Fig. 28).
- 23-