R-4108.A

 

REPAIR ANO AOJUSTMENT

1. Failure of the Slide Glass Retainer 2. Slack of the Slide Glass Carrier

 

3. Wobbling of the Slide Glass Retainer and the Slide Glass Carrier

 

 

 

 

 

 

REPAIR ANO AOJUSTMENT

1.Failure of the Slide Glass Retainer

Cause: (1) The yielded or disengaged sp1ral spring @.

(2)Looseness of the small t1at head screw@.

<Repair and Adjustment >

(!)Exchange the yielded spiral spring @ or hook the disengaged spring to the pin hole.

Procedure j

'··,,

i) Disassemble the retamer assembly following the procedure 2), 3) (P. 5).

ü) If the spring is yielded replace it with a spare, and if the spring is disengaged, reposition it correctly foilowing the reassembling proced.ure (P. 7).

ili) Check the function of repaired retainer.

(2)If the small ílat head screw @ is loosened fasten it firm and check the retainer function.

2.Slack of the Slide Gim Carrier

Cause: Looseness of the screws @

<Repair and Adjustment >

Fasten firm the screws @ . Make sure that no slack is found.

3.Wobbling of the Slide Glass Retainer and the Slide Gim Carrier.

Wobbling in X direction: Refer to the following chapter 4.

Wobbling in Y direction: Refer to the following chapter 5.

·~)

- 12 -

Page 61
Image 61
Nikon 50i and E100 Failure of the Slide Glass Retainer, Repair and Adjustment, Procedure j, Slack of the Slide Gim Carrier