Appendix G: Product Verification
Appendices
A --- 2 4
HSet the WAVELENGTH(nm)display to read 1300.
HPush the WAVELENGTH button once more; check that the button is
no longer backlitand that the MODE button is backlit. (After exiting
the wavelength mode, the instrument may require a secondor two
to recalculate calibrationfor the new wavelength; the ATT ADJ light
next to the ATTENUATION(dB) display will brief ly be backlit to indi-
cate this.)
Step 1: Push the ADDR(SET)button once.
HThe button should becomeback lit,the ATTENUATION display
should read Addr,and the WAVELENGTH display should show a
number from 0to 30 or the word OFF.
Step 2: While in thismode, the coarse and fine knobs should increment
or decrement the displayedaddress value by 1; adjust the value, using
the knobs, to the address desired.
HCheck that when the value is adjusted above 30that the word
OFFappears.
Step 3: Push the ADDR(SET)button once again while the desired
address is being displayed;the display mode should return to whatever
state the instrument was in before pushing the ADDR(SET)button.
NOTE
The address of the instrument is not actually updated until the
ADDR (SET) mode is exited.
Step 1: Enter the Wavelength set modefor the OA 5000 by pushing the
WAVELENGTH button so that the display reads SEt.
Step 2: Using the knobs, adjust the wavelength to read 1350 nm.
Step 3: Exit the WavelengthSet mode by pushing the WAVELE NGTH
button again; the instrumentw illrequire a second to prepare the new
calibration tables (as indicated by the ATT AD J briefly lighting up).
Step 4: Adjust the OA 5000 such that the display modeis showing
attenuation in dB with the ATTindicator lit and the ---REF indicator not lit.
Step 5: Using the knobs, increase the level ofattenuation (turn the
coarse knob clockwise); keep increasingthe level until the value reach-
es a maximum level and nolonger increases. This value should exceed
60.00 dB.
GPIB Address SettingAttenuation Range