OA 5000 Series User Manual 2 --- 3
Enabling/Disabling Attenuation
On the OA 5000, the DISABLEbutton (Figure 2-2) allows you to enable or
disable the opticalpath through the attenuator with a shutter. You can still
set the attenuation level, however,since it is not dependent on the optical
signal passing throughthe attenuator. For example, you can block the
passage of the opticalsignal through the attenuator by disabling attenua-
tion, setting the attenuationlevel, then enabling attenuation and your signal
will be attenuated to the desired level immediately.
DISABLE
MINATT
STORE1 RECALL1
STORE2 RECALL2
Thisbutton enables or
disablesattenuation.
Figure 2-2: The Location of the DISABLE Button
Toenable attenuation, press the DISABLE button on the front panel so that
the button is not lit. When the DISABLEbutton is not lit, the optical signal
passes through the attenuator.
Todisable attenuation, press the DISABLE button so that it is lighted. When
the DISABLEbutton is lighted, the internal shutter is placed across the
optical path, providing 100 dB attenuationfor the OA 5002, OA 5012, and
OA 5022, and 90 dB attenuation for the OA5032.

WARNING

Topr event damage to your eyes, avoid looki ng into the optical
output port while there i s an optical signal con nected to the input
port. Thedisable function is only valid while the instrumentis
switched on. When theOA 5000 is switched off, the shutter with-
draws from the optical path and the light is attenuated only by the
previous attenuator setting.