GPIB RequirementsObserve these rules when using yourOA 5000 with a GPIB network:
HEach device on the bus must be assigned a uniquedevice address; no
two devices can share the same device address.
HDo not connect more than 15 devicesto any one bus.
HConnect one device forevery 6 feet (2 meters) of cable used.
HDo not use more than 65 feet (20 meters) of cableto connect devices to
abus.
HAt least two-thirds of the deviceson the network must be turned on
while the network is operating.
HConnect the devices on the network in a star orlinear configuration as
shown in Figure 3-2. Do not use loop or parallel con figurations.
GPIBDevice
GPIBDevice
GPIBDevice
GPIBDevice
GPIBDevice
GPIBDevice
GPIBDevice
Figure 3-2: TypicalG PIB Network Configurations