OA 5000 Series User Manual 2 --- 5
Setting Attenuation LevelsBecause the OA 5000 automaticallyadjusts the attenuator to correct for
different wavelengths, it isimportant, for attenuator accuracy, to correctly
match the displayed wavelength with the wavelength ofthe signal going into
the unit.
Setting attenuation levelson the OA 5000 requires you to perform two
procedures in the followingorder:
1. Specify the wavelength of the signalto be attenuated.
2. Set the level of attenuation.
Refer to Figure 2-3.
DISABLE
MINATT
STORE1 RECALL1
STORE2 RECALL2
REMOTE MODE
SETREF COARSE
FINE
WAVE-
LENGTH
ADDR
(SET)
Thisbutton sets
attenuationto the
minimum
value.
Theseknobs set the
attenuationlevel,
referencelevel,
wavelengthand GPIB
address.
Thisbutton assigns the
ATTENUATIONdisplay to
eitherabsolute units or
units relative to the
referencelevel.
Usethis button to assign the
knobsto adjust the
wavelength.
Figure 2-3: Setting Attenuation Levels
Tospecify the wavelength to be attenuated:
Step 1: Pressthe WAVELENGT H button.
The word SEt will appear inthe attenuation display and the WAVE-
LENGTH button will illuminate.