
OA 5000 Series User Manual A --- 3
Appendix B: SpecificationsThe performance characteristics describedin this appendix (Tables A-1
through A-3) are valid under the followingconditions:
HThe instrument must have been calibrated at an ambienttemperature
between +20_C and +30_C.
HThe instrument must have had a warm-up period of at least20 minutes.
HAll opticalconnections to the instrument are made using physical con-
tact connectors. Tektronix offers limite d selection of fiber optic jumper
cables which utilize physical contact on t he connector ends (for exam-
ple: Tektronixpart number 174-1387-00, see Appendix A).
HBoth input and output connectorshave been cleaned as described on
page A---15.
TableA-1: Optical Characteristics
Characteristic Specification
Wavelength Range OA 5002 1100 ---1600 nm
OA 5012/22/32 750 ---1600 nm
Input
OA 5002
OA 5012
OA 5022
OA 5032
Single Mode Fiber
50 m Multimode Fiber
62.5 m Multimode Fiber
100 m Multimode Fiber
Insertion Loss
750 ---1100 nm
1100---1350nm
1350---1600nm
≤4.0 dB (OA 5012/22/32 only)
≤2.0 dB
≤2.5 dB
Return Loss
OA 5002
OA 5012/5022/5032
≤---4 5 d B
≤---2 5 d B
Attenuation Range
OA 5002
1100 ---1350 nm
1350---1600nm
OA 5012/22/32
750---1350nm
1350---1600nm
≥60 dB (calibrated, 0.01 dB step size)
≥50 dB (calibrated, 0.01 dB step size)
≥60 dB (calibrated, 0.01 dB step size)
≥50 dB (calibrated, 0.01 dB step size)
Maximum Attenuation
OA 5002/5012/5022
OA 5032
≥100 dB (shutter activated)
≥90 dB (shutter activated)
Repeatability ±0.05dB