OA 5000 Series User Manual A --- 3
Appendix B: Specifications
The performance characteristics describedin this appendix (Tables A-1
through A-3) are valid under the followingconditions:
HThe instrument must have been calibrated at an ambienttemperature
between +20_C and +30_C.
HThe instrument must have had a warm-up period of at least20 minutes.
HAll opticalconnections to the instrument are made using physical con-
tact connectors. Tektronix offers limite d selection of fiber optic jumper
cables which utilize physical contact on t he connector ends (for exam-
ple: Tektronixpart number 174-1387-00, see Appendix A).
HBoth input and output connectorshave been cleaned as described on
page A---15.
TableA-1: Optical Characteristics
Characteristic Specification
Wavelength Range OA 5002 1100 ---1600 nm
OA 5012/22/32 750 ---1600 nm
Input
OA 5002
OA 5012
OA 5022
OA 5032
Single Mode Fiber
50 m Multimode Fiber
62.5 m Multimode Fiber
100 m Multimode Fiber
Insertion Loss
750 ---1100 nm
1100---1350nm
1350---1600nm
4.0 dB (OA 5012/22/32 only)
2.0 dB
2.5 dB
Return Loss
OA 5002
OA 5012/5022/5032
---4 5 d B
---2 5 d B
Attenuation Range
OA 5002
1100 ---1350 nm
1350---1600nm
OA 5012/22/32
750---1350nm
1350---1600nm
60 dB (calibrated, 0.01 dB step size)
50 dB (calibrated, 0.01 dB step size)
60 dB (calibrated, 0.01 dB step size)
50 dB (calibrated, 0.01 dB step size)
Maximum Attenuation
OA 5002/5012/5022
OA 5032
100 dB (shutter activated)
90 dB (shutter activated)
Repeatability ±0.05dB