Commands
OA 5000 Series User Manual 3 --- 2 5
Table3-6: FACTORYFront Panel Settings (Cont.)
Front Panel Parameter Setting
DISABLE OFF
WAVELENGTH 1300
The FACTORYcommand does not alter the following items:
HThe state of the GPIB (IEEE Std 488.2) interfaces.
HThe selected GPIB address.
HCalibrationdata that affects device specifications.
HEADer
The HEADer command sets and queries the Response HeaderEnable State
that causes the OA 5000 to either includeor omit headers on query re-
sponses. This command does notaffect IEEE Std 488. 2 CommonCom-
mands (those starting with an asterisk) or the *LRN?response.
VERBOSE.
HEADer { ON | OFF | <NRf> }
HEADer?
<NRf>
?
<Space>
ON
OFF
HEADer
ON or <NRf> ΒΈ0 sets the Response Header Enable State toTRUE . This
causes the OA 5000 to includeheaders on applicable query responses. You
can then use the query response as a command.
OFF or <NRf> = 0 sets the Response Header Enable State toFALSE. This
causes the OA 5000 to omitheaders on query responses, so that only the
argument is returned.
HEADER OFF
causes the OA 5000 to omitheaders from query responses.
HEADER 1
causes the OA 5000 to includeheaders on applicable query responses.
Related Commands:
Syntax:
Arguments:
Examples: