Commands
Programming
3 --- 3 0
*RST does the following:
HPuts the OA 5000 intothe Operation Complete Command Idle State.
HPuts the OA 5000 intothe Operation Complete Query Idle State.
HReturns the instrument settings to thoselisted on page 3 ---24.
The *RST command does not alterthe following:
HThe state of the IEEE Std 488.1 interface.
HThe selected IEEE Std 488.1 address of the OA 5000.
HCalibrationdata that affects device specifications.
HThe Output Queue.
HThe Standard Status Register Enable setting.
HThe Standard Event Status Enable setting.
HThe Power-onstatus clear flag setting.
HHEADer and VERBose settings.
*SRE
The *SRE (Service Request Enable) command sets and queriesthe bits in
the Service Request Enable Register (SRER). Fora complete discussion of
the use of these registers, see page 3---35.
ALLev?, *CLS, DESE, *ESE, *ESR, EVENT?, EVMSG?, EVQty?, *PSC,
*STB.
*SRE <NRf>
*SRE?
?
*SRE
<Space> <NRf>
<NRf> is a value in the range from0 to 255. The binary bits of the SRER are
set according to thisvalue. Using an out-of-range value causes an execution
error.
The power-ondefault for SRER is 0 if *PSC is 1. If *PSC is 0, the SRER
maintains its valuethrough a power cycle.
Related Commands:
Syntax:
Arguments: