Setting the Reference Level
OA 5000 Series User Manual 2 --- 1 3
The second applicationfor reference mode is approximating signal power
after the source has been attenuated.
Step 1: Set the attenuator to its minimumreference level.
Step 2: Measure the power (in dBm) from the fiber connectedto the
output optical connector.
Step 3: Set the reference value to this number.
Once this levelhas been set and the OA 5000 is in reference mode, the
displayed attenuationis the negative of the power level in dBm. For exam-
ple, if the power level is ---2.5dBm when the attenuator is at its minimum
attenuation leveland this is entered as the reference level, then the display
will read 2.5 dB when set to REF mode. Then, if youincrease the attenuation
by 10 dB, the display willread 12.5, which is the negativeof the power level
in dBm.
Approximating SignalPower