Commands
Programming
3 --- 3 4
*WAI (No Query Form)
The *WAI (Wait)command prevents the OA 5000 from executing further
commands or queries untilall pending operations finish.
*OPC, ADJusting?.
*WAI
*WAI
WAVelength
This commands set or queriesthe w avelengthused by the instrument when
determining the equivalentattenuator positioning for a given attenuation.
N/A
WAVelength <NRf>
WAVelength?
WAVelength
<Space>
?
<NRf>
The valid range for the wavelength argument is600 nm to 1700 nm. D efault
units are always in nanometersw ithN (for nano or 1E ---09) and U (for micro
or 1E---06)accepted as multipliers; M (for meters) is also an acceptable term
(see examples below).
WAV?
returns the value of wavelength to which the OA5000 is set. An example
response is:
:WAVELENGTH1300
WAV 1300
sets the wavelength to 1300 nm.
WAV 1300NM
sets the wavelength to 1300 nm.
WAV 1.3UM
sets the wavelength to 1300 nm.
WAV 1.3E-09M
sets the wavelength to 1300 nm.
Related Commands:
Syntax:
Related Commands:
Syntax:
Arguments:
Examples: