Calibrating with the X-Rite DTP34 QuickCal Densitometer

115

Calibrating the X-Rite DTP34 QuickCal Densitometer

Frequency of Calibration

Your X-Rite DTP34 QuickCal densitometer should have a quick calibration at least once a day. A full calibration should be performed annually, during heavy usage to provide accurate measurements, or if the quick calibration consistently fails. Typically, the host computer prompts for an instrument calibration (amber LED) when required, but the process can be invoked manually at any time. See the following procedure for information about the calibration process.

Positioning the Instrument on the Calibration Reference

Handle the reflection reference by the edges. Make sure that the reflection reference is free of dust, dirt, and smudge marks. To obtain the most accurate calibration, hold the instrument with consistent and nominal pressure during the calibration process.

Position the instrument on the designated location of the reflection reference (indicated by a dotted outline of the instrument). Do not move the instrument more than 0.25" (6.35mm) before reading the strip.

DTP34-100 #: AXXXXXXXX-XXXXX D: XXX

STATUS X

C= .XX M= .XX Y= .XX

X-RiteR

Reflection Reference

Page 125
Image 125
Xerox CX250, CX240 manual Calibrating the X-Rite DTP34 QuickCal Densitometer