•Press the [DEFINE CONSTANT] softkey (F2).
•Press the [DEFINE K1] softkey (F1).
•Enter 386.4 and press the [ENTER] softkey (F1).
•Press the [Analys] hardkey.
•Press the [DEFINE FUNCTION] softkey (F1).
•Press the [DEFINE F1] softkey (F1).
•Press the [CONSTANT
•Press the [CONSTANT K1] softkey (F1).
•Press the [ * ] softkey (F3).
•Press the [OPERATION] softkey (F5).
•Press the [MORE] softkey (F9).
•Press the [ /jOMEGA( ] softkey (F4).
•Press the [OPERATION] softkey (F5).
•Press the [MORE] softkey (F9).
•Press the [ /jOMEGA( ] softkey (F4).
•Press the [MEAS DATA] softkey (F1).
• Press the [FREQ RESP 2 / 1] softkey (F5).
•Press the [ ) ] softkey (F5).
•Press the [ ) ] softkey (F5).
•Press the [ENTER] softkey (F7).
Now display the function F1 on Trace B
•Press the [Active Trace] hardkey.
•Press the [B] softkey (F2).
•Press the [Meas Data] hardkey.
•Press the [MORE CHOICES] softkey (F9).
•Press the [MATH FUNCTION] softkey (F7).
• Press the [PWR SPEC CHANNEL 1] softkey (F3)
•Press the [F1] softkey (F1).
Trace B will now convert “g/lb.” into “inch/lb.” This same technique could also be used to convert “g” into velocity terms, or to convert between metric and English units.