Trigger Delay

AGILENT 35670A

Supplemental Operator’s Guide

Then set up trigger parameters (for a positive event, positive slope) as follows:

Press the [TRIGGER SETUP] softkey (F6)

Press the [CHANNEL LEVEL] softkey (F1)

Enter 5 then press the [PERCENT (%)] softkey (F4)

Press the [SLOPE POS NEG] softkey (F5) until POS is highlighted.

Trigger Delay

Both channels should acquire data using the same pre-trigger delay. To set up a

0.01s pre-trigger condition, follow these steps:

Press the [Trigger] hardkey.

Press the [TRIGGER SETUP] softkey (F6)

Press the [ALL CHANNELS] softkey (F8)

Press the [CHANNEL * DELAY] softkey (F9)

Enter -.01then press the [S] softkey (F1)

To start the data acquisition,

Press the yellow [Start] hardkey.

The analyzer should show in highlighted text above the trace:

WAITING FOR CH 1 TRIGGER

Trigger

Arm: Automatic

Waterfall Steps: 15

[FFT]

Chn Level: 5 %

Slope: Positive

Date: 10-03-96 Time: 05:23:00 PM

 

A: CH1 Time

X:472.168 ms

Y:371.577 mV

500

 

 

mV

 

 

Real

 

 

100

 

 

mV

 

 

/div

 

 

-500

 

 

mV

 

 

-9.7656ms

489.7461ms

B: CH2 Time

X:472.168 ms

Y:-241.753 mV

500

 

 

mV

 

 

Real

 

 

100

 

 

mV

 

 

/div

 

 

-500

 

 

mV

 

 

-9.7656ms

489.7461ms