significantly. Ten averages may take longer, but may be necessary when testing difficult, non-linear structures. To set up five averages, follow these steps:

Press the [Avg] hardkey.

Toggle the [AVERAGE ON OFF] softkey (F1) until ON is highlighted.

Press the [NUMBER AVERAGES] softkey (F2).

Enter 5 and press the [ENTER] softkey (F1).

To avoid averaging overloaded data, turn on overload reject as follows:

Press the [Avg] hardkey.

Toggle the [OVLD REJ ON OFF] softkey (F1) until ON is highlighted.

To start taking measurements,

Press the yellow [Start] hardkey.

Hammer the structure at one measurement point. After the data is acquired and displayed, hammer the structure again, and repeat 5 times until the average is complete.

Avg

Type: RMS

Number: 5

[FFT]

Update Rt: 5

Overlap: 0 %

Date: 10-02-96 Time: 09:06:00 PM

 

 

 

A: Freq Resp2/1

X:2.96

kHz

 

 

Y:98.678

100

 

 

 

 

 

LogMag

 

 

 

 

 

4

 

 

 

 

 

decades

 

 

 

 

 

10

 

 

 

 

 

e-3

 

 

AVG:

5

3.2kHz

0Hz

 

 

B: Coherence2/1

X:2.96

kHz

 

 

Y:963.738 e-3

1

 

 

 

 

 

Real

 

 

 

 

 

100

 

 

 

 

 

e-3

 

 

 

 

 

/div

 

 

 

 

 

0

 

 

 

 

 

0Hz

 

 

AVG:

5

3.2kHz

30

Page 34
Image 34