Scaling for traces A and B are now identical. To accomplish the same trace scaling for C and D, follow these steps:

Press the [Active Trace] hardkey.

Press the [C D] softkey (F1).

Press the [Scale] hardkey.

Toggle the [AUTOSCALE ON OFF] softkey (F1) until ON is highlighted.

Press the [Y PER DIV (DECADES)] softkey (F6).

Enter 2 and press the [ENTER] softkey (F1).

As with Traces A and B, if the top reference of C and D are not identical, set the top reference level as shown above.

Using Markers to Compare

Coupled markers can be activated for all traces, but not independently for each pair of traces. Use coupled markers to compare one set of traces at a time.