Set up the AGILENT 35670A

AGILENT 35670A

Supplemental Operator’s Guide

difference in motion between Channel 1 and Channel 2 in mils peak-to- peak

phase angle between the two channels.

Set up the AGILENT 35670A

Press the green [Preset] hardkey

Press [DO PRESET] softkey (F1).

Press the [Pause/Cont] hardkey if desired.

Press the [Freq] hardkey

Enter 800 and then press the [Hz] softkey (F2).

Press the [Input] hardkey.

Press the [ALL CHANNELS] softkey (F2).

Select the [CH* AUTO UP ONLY] softkey (F5).

Set up transducer parameters on Channel 1 and Channel 2 as per the section “Measuring with Transducers.” Mount the accelerometers to the machinery and connect them to Channel 1 and Channel 2. The following procedures assume that

Using a Math Function to Measure Differential Motion

For accelerometers with the same polarity

Press the [Analys] hardkey.

Press the [DEFINE FUNCTION] softkey (F1).

Press the [DEFINE F1] softkey (F1).

Press the [MEAS DATA] softkey (F1).

Press the [LIN SPEC CHANNEL 1] softkey (F3).

Press the [ - ] softkey (F2).

Press the [MEAS DATA] softkey (F1).

Toggle the [CHANNEL 1 2 3 4] softkey (F1) until 2 is highlighted.

Press the [LIN SPEC CHANNEL 2] softkey (F3).

Press the [ENTER] softkey (F7).

For accelerometers with opposite polarity

Press the [Analys] hardkey.

Press the [DEFINE FUNCTION] softkey (F1).

Press the [DEFINE F1] softkey (F1).

Press the [MEAS DATA] softkey (F1).

Press the [LIN SPEC CHANNEL 1] softkey (F3).

Press the [ + ] softkey (F1).

Press the [MEAS DATA] softkey (F1).