Quantifying Time Trace Results

AGILENT 35670A

Supplemental Operator’s Guide

Input

CH1 Rng: 1.0012 Vpk

CH2

Rng: 631.706 mVpk

 

CH3 Rng: 2.5149 Vpk

CH4

Rng: 2.5149 Vpk

Date: 10-02-96 Time: 08:47:00 PM

 

A: CH1 Time

X:38.0859 ms

Y:1.06837 V

2.5

 

 

V

 

 

Real

 

 

500

 

 

mV

 

/div

 

 

-2.5

 

 

V

-9.7656ms

489.7461ms

 

Quantifying Time Trace Results

Use the marker knob to read the amplitude at any time during the trace. Rotate the knob on the AGILENT 35670A front panel until the marker indicator (the small diamond following the spectral trace) is placed at the desired time. The time and amplitude at this bin is presented as X and Y values at the top of the trace. Units of [s] and [Vrms] or [grms] identify the results.

Peak tracking is very useful for determining maximum amplitude of positive peaks in the time domain:

Press the [Marker] hardkey.

Press [PEAK TRK ON OFF] softkey (F6) until ON is highlighted.

Care must be taken with negative peaks, however, because PEAK TRK ON only looks for the maximum value, and large negative peaks will be ignored.

14