AGILENT 35670A
Supplemental Operator’s Guide
Section 13 : Order Domain Results in List Mode
Machinery
To set up and measure a list of 10 orders for a machinery
•Press the green [Preset] hardkey
•Press [DO PRESET] softkey (F1).
•Press the [Pause/Cont] hardkey if desired.
•Press the [ORDER ANALYSIS] softkey (F3).
•Press [CHANNELS 1 2 4 ] softkey (F7) until 1 is highlighted.
•Press the [Input] hardkey
•Press the [CH1 AUTO UP ONLY] software (F5).
Order analysis requires a tachometer signal connected to the TACH BNC at the back of the analyzer. To set up the AGILENT 35670A for the tachometer signal, follow these steps:
•Press the [TACHOMETR SETUP] softkey (F9).
• Toggle the [TACH DISP ON OFF] softkey (F6) until ON is highlighted.
•Press the [TACH PULS PER REV] softkey (F1).
•Enter the number of tachometer pulses per shaft revolution and press the [ENTER] softkey (F1).
•Press the [LEVEL] softkey (F3).
•Enter the tachometer level and press the [ENTER] softkey (F1).
If in doubt about the tachometer level, enter 0.5 V and experiment with level, slope, and trigger range until a tachometer display is shown.
Prepare the order display as follows:
•Press the [Disp Format] hardkey.
•Press the [UPPER/LOWER] softkey (F3).
•Press the [Active Trace] hardkey.
•Press the [B] softkey (F1)
Prepare the order results list as follows:
•Press the [Analys] hardkey.
•Press the [DATA TABLE] softkey (F5).
•Toggle the [TABLE ON OFF] softkey (F1) until ON is highlighted.
•Press the [INSERT X VALUE] softkey (F4).
•Enter 1 and press the [ORDERS] softkey (F1).
•Enter 2 and press the [ORDERS] softkey (F1).
•Enter 3 and press the [ORDERS] softkey (F1).
33