Section 6 : Measuring a Single Channel Time Waveform

This section contains procedures for setting up and measuring a single channel time waveform using an AGILENT 35670A Dynamic Signal Analyzer. Included are steps to select single channel operation, set up the trace time length, and set up triggering to capture a transient event. Instructions will also be presented to scale the time trace.

Selecting Single Channel Operation

To measure a single-channel time waveform, turn on the AGILENT 35670A and then:

Press the [Inst Mode] hardkey

Press [CHANNELS 1 2 4 ] softkey (F7) until 1 is highlighted.

Note: if the AGILENT 35670A was already turned on and in an unknown state, then preset the analyzer by doing the following:

Press the green [Preset] hardkey

Press [DO PRESET] softkey (F1).

Then set the instrument mode to single channel as above.

Time Record Length in Seconds vs. Resolution

Span - Hz

100 lines

200 lines

400 lines

800 lines

1600 lines

51200 Hz

0.001953 s

0.003906 s

0.007813 s

0.015625 s

0.03125 s

25600 Hz

0.003906 s

0.007813 s

0.015625 s

0.03125 s

0.0625 s

12800 Hz

0.007813 s

0.015625 s

0.03125 s

0.0625 s

0.125 s

6400

Hz

0.015625 s

0.03125 s

0.0625 s

0.125 s

0.25 s

3200

Hz

0.03125 s

0.0625 s

0.125 s

0.25 s

0.5 s

1600

Hz

0.0625 s

0.125 s

0.25 s

0.5 s

1 s

800 Hz

0.125 s

0.25 s

0.5 s

1 s

2 s

400 Hz

0.25 s

0.5 s

1 s

2 s

4 s

200 Hz

0.5 s

1 s

2 s

4 s

8 s

100 Hz

1 s

2 s

4 s

8 s

16 s

50 Hz

2 s

4 s

8 s

16 s

32 s

25 Hz

4 s

8 s

16 s

32 s

64 s

12.5

Hz

8 s

16 s

32 s

64 s

128 s

6.25

Hz

16 s

32 s

64 s

128 s

256 s

3.125

Hz

32 s

64 s

128 s

256 s

512 s

1.5625

Hz

64 s

128 s

256 s

512 s

1024 s

0.78125

Hz

128 s

256 s

512 s

1024 s

2048 s

0.390625

Hz

256 s

512 s

1024 s

2048 s

4096 s

0.1953125

Hz

512 s

1024 s

2048 s

4096 s

8192 s