WAITING FOR CH 1 TRIGGER

AGILENT 35670A

Supplemental Operator’s Guide

Press the [TRIGGER SETUP] softkey (F6)

Press the [ALL CHANNELS] softkey (F8)

Press the [CHANNEL * DELAY] softkey (F9)

Enter -.01then press the [S] softkey (F1)

Now to check that the trigger parameters are correct,

Press the yellow [Start] hardkey.

The analyzer should show in highlighted text above the trace:

WAITING FOR CH 1 TRIGGER

Tap the hammer on the test structure to see if the measurement will trigger. If it will not, adjust the trigger level and slope until successful.

Setting Up Time Displays

View both the time trace and the windowed time trace for both channels using the QUAD display format as follows:

Press the [Disp Format] hardkey.

Press the [QUAD] softkey (F6).

Press the [Active Trace] hardkey.

Press the [A] softkey (F1)

Press the [Meas Data] hardkey.

Press the [CHANNEL 1 2 3 4] softkey (F1) until 1 is highlighted.

Press the [TIME CHANNEL 1] softkey (F5)

Press the [Active Trace] hardkey.

Press the [B] softkey (F1)

Press the [TIME CHANNEL 2] softkey (F5)

Press the [Active Trace] hardkey.

Press the [C] softkey (F1)

Press the [CHANNEL 1 2 3 4] softkey (F1) until 1 is highlighted.

Press the [MORE CHOICES] softkey (F9)

Press the [WINDOWED TIME CH 1] softkey (F5)

Press the [Active Trace] hardkey.

Press the [D] softkey (F1)

Press the [WINDOWED TIME CH 2] softkey (F5)

Press the [Active Trace] hardkey.

Press the [A B C D] softkey (F7).

Press the [Scale] hardkey.

Toggle the [AUTOSCALE ON OFF] softkey (F1) until ON is highlighted.

27