Agilent Technologies Agilent 35670A Selecting Single Channel Operation, Selecting Frequency Span

Models: Agilent 35670A

1 82
Download 82 pages 27.58 Kb
Page 9
Image 9
Selecting Single Channel Operation

AGILENT 35670A

Supplemental Operator’s Guide

Section 4 : Measuring a Single Channel Power Spectrum - Displacement

This section contains procedures for setting up and measuring a single channel power spectrum of displacement using an AGILENT 35670A Dynamic Signal Analyzer. Included are steps to select single channel operation, set up transducer parameters, frequency span, choose input range, select windowing, and turn on averaging in order to improve measurement accuracy. Instructions will also be presented to set up trace coordinates and trace scaling.

Selecting Single Channel Operation

Specify the single-channel instrument mode to measure one signal. However it should be noted that it is acceptable to measure a single channel in the default two- channel instrument mode and simply ignore the second channel. Nevertheless instructions will be given here to measure in single channel mode.

To measure a single-channel spectrum, turn on the AGILENT 35670A and then:

Press the [Inst Mode] hardkey

Press [CHANNELS 1 2 4 ] softkey (F7) until 1 is highlighted.

Setting Up Transducer Units for Displacement

To measure power spectrums using a displacement transducer with a sensitivity of

200 mV/mil, follow these steps:

Press the [Input] hardkey.

Enter 800 and then press the [Hz] softkey (F2).

Any frequency below 51.2 kHz may of course be entered, however the analyzer will select the lowest span that contains the desired frequency. For example, if 1200 Hz is entered, the analyzer is switched to a span of 1600 Hz.

Selecting Frequency Span

The default frequency span after turn-on is 51.2 kHz, which is usually too high for most transducer measurements. To reduce the span to 800 Hz, perform the following steps:

Press the [Freq] hardkey.

Enter 800 and then press the [Hz] softkey (F2).

5

Page 9
Image 9
Agilent Technologies Agilent 35670A manual Selecting Single Channel Operation, Setting Up Transducer Units for Displacement