AGILENT 35670A
Supplemental Operator’s Guide
Section 4 : Measuring a Single Channel Power Spectrum - Displacement
This section contains procedures for setting up and measuring a single channel power spectrum of displacement using an AGILENT 35670A Dynamic Signal Analyzer. Included are steps to select single channel operation, set up transducer parameters, frequency span, choose input range, select windowing, and turn on averaging in order to improve measurement accuracy. Instructions will also be presented to set up trace coordinates and trace scaling.
Selecting Single Channel Operation
Specify the
To measure a
•Press the [Inst Mode] hardkey
•Press [CHANNELS 1 2 4 ] softkey (F7) until 1 is highlighted.
Setting Up Transducer Units for Displacement
To measure power spectrums using a displacement transducer with a sensitivity of
200 mV/mil, follow these steps:
•Press the [Input] hardkey.
•Enter 800 and then press the [Hz] softkey (F2).
Any frequency below 51.2 kHz may of course be entered, however the analyzer will select the lowest span that contains the desired frequency. For example, if 1200 Hz is entered, the analyzer is switched to a span of 1600 Hz.
Selecting Frequency Span
The default frequency span after
•Press the [Freq] hardkey.
•Enter 800 and then press the [Hz] softkey (F2).
5